Fluke 19xC-2x5C
Users Manual
130
Meter Mode (Vdc, Vac, Vac+dc, Ohm and Continuity): Reading disturbance with test leads shorted
Table 15
Disturbance less than 1% of full scale
E = 3V/m
Frequency range 10 kHz to 1 GHz
500 mV to 1000 V , 500 Ohm to 30 MOhm ranges
Table 16
No visible disturbance
E = 3V/m
Frequency range 1.4 GHz to 2 GHz
500 mV to 1000 V, 500 Ohm to 30 MOhm ranges
Table 17
No visible disturbance
E = 1V/m
Frequency range 2 GHz to 2.7 GHz
500 mV to 1000 V , 500 Ohm to 30 MOhm ranges