Vital Signs Simulator
Detailed Specifications
47
Temperature
Temperature
.............................................................. 0
°
C (32
°
F), 24
°
C (75.2
°
F), 37
°
C (98.6
°
F), and 40
°
C (104
°
F)
Accuracy
....................................................................
±
0.1
°
C
Compatibility
............................................................. Yellow Springs, Inc. (YSI) Series 400 and 700
Output Connector
..................................................... DIN 4-pin
Power-On Default
...................................................... 0
°
C (42
°
F)
Cardiac Output (ProSim 3 Only)
Catheter Type
............................................................ Baxter Edwards, 93a-131-7f
Calibration Coefficient
............................................. 0.542 (0
°
C injectate), 0.595 (24
°
C injectate)
Blood Temperature
................................................... 37
°
C (98.6
°
F)
±
2 %
Injectate Volume
....................................................... 10 cc
Injectate Temperature
.............................................. 0
°
C or 24
°
C
±
2 % value
Cardiac Output
.......................................................... 2.5, 5, 10 liters per minute
±
5 %
Faulty-Injectate Curve
.............................................. Waveform for simulation available
Left-to-Right-Shunt Curve
........................................ Waveform for simulation available
Calibrated Pulse
........................................................ 1.5
°
for 1 second (37
°
to 35.5
°
)
Output Connector
..................................................... DIN 7-Pin
Power-On Default
..................................................... 2.5 liters per minute, 0
°
C injectate
Fetal / Maternal-ECG (ProSim 3 Only)
Fetal Heart Rate (Fixed)
............................................ 60, 90, 120, 140, 150, 210, and 240 BPM
Fetal Heart Rate (IUP)
: .............................................. 140 BPM at beginning, then varies with pressure
Intrauterine-Pressure Waveforms
........................... Early deceleration, late deceleration, and uniform acceleration
Wave Duration
........................................................... 90 seconds, bell-shaped pressure curve, from 0 mmHg to 90 mmHg and returning to 0 mmHg
IUP Period
.................................................................. 2, 3, or 5 minutes, and manual
Power-On Default
...................................................... FHR 120 BPM, early deceleration, manual