Vital Signs Simulator
General Specifications
43
General Specifications
Power
......................................................................... Two 9-V alkaline batteries (IEC 6LR61, NEDA 1604A). Optional battery eliminator: 15 Vdc, 1.5 mA
Battery Life
................................................................ 8 hours minimum
Display
....................................................................... LCD Greyscale Display
Size
............................................................................. 14.0 cm x 20.6 cm x 4.5 cm (5.5 in x 8.2 in x 1.8 in)
Weight
........................................................................ 0.47 kg (1 lb 4 oz)
Temperature
Storage ................................................................ -25
°
C to +50
°
C (-13
°
F to +122
°
F)
Operation ............................................................. 10
°
C to 40
°
C (50
°
F to 104
°
F)
Humidity
.................................................................... 10 % to 80 % non-condensing
Altitude
...................................................................... 2000 m (6,562 ft)
Safety
......................................................................... IEC 61010-1, Pollution degree 2
Electromagnetic Environment
................................. IEC 61326-1, Portable
EMC
............................................................................ Applies to use in Korea only. Class A equipment (Industrial Broadcasting & Communication Equipment)
[1]
[1] This Product meets requirements for industrial (Class A) electromagnetic wave equipment and the
seller or user should take notice of it. This equipment is intended for use in business environments and
is not to be used in homes.
Detailed Specifications
ECG Waveform
ECG Reference
.......................................................... The ECG amplitudes specified are for Lead II, from the baseline to the peak of the R wave. All other
leads are proportional in percentage per:
Lead I .................................................................. 70 %
Lead II ................................................................. 100 %
Lead III ................................................................ 30 %
Lead V1 ............................................................... 24 %
Lead V2 ............................................................... 48 %
Lead V3 ............................................................... 100 %
Lead V4 ............................................................... 120 %
Lead V5 ............................................................... 112 %