Vital Signs Simulator
Detailed Specifications
45
Ventricular Arrhythmia
............................................. PVCs (6, 12, or 24 per minute), frequent multifocal PVCs, bigeminy, trigeminy, multiple PVCs (one-time
run of 2, 5, or 11 PVCs), ventricular tachycardia, ventricular fibrillation (coarse or fine), and asystole
Conduction Defect
.................................................... First-, second-, or third-degree AV block and right- or left-bundle-branch block
Power-On Default
...................................................... None (off)
ECG-Performance-Tests
Amplitude
.................................................................. 0.05 mV to 0.45 mV (0.05 mV steps)
0.5 mV to 5.5 mV (0.5 mV steps)
Pulse Wave
................................................................ 30 BPM, 60 BPM, with 60 ms pulse width
Square Wave
............................................................. 2.0, 0.125 Hz
Triangle Wave
........................................................... 2.0, 2.5 Hz
Sine Wave
.................................................................. 0.5, 5, 10, 40, 50, 60, 100 Hz
R-wave-Detection Waveform
................................... Haver-Triangle
R-wave Rate
.............................................................. 30, 60, 80, 120, 200, and 250 BPM
R-wave Width
............................................................ 20 ms to 200 ms (10 ms steps)
Additional Steps: 8, 10, and 12 ms
Rate Accuracy
...........................................................
±
1 %
Amplitude Accuracy
.................................................
±
2 %, Lead II (Exception:
±
5 % for R waves
≤
20 ms)
Power-On Default
..................................................... None (off)
Respiration
Rate
............................................................................ 0 (OFF), 15, 20, 30, 40, 60, 80, 100, 120 BrPM
Impedance Variations (
Δ
Ω
)
..................................... 0.2, 0.5, 1, or 3
Ω
peak-to-peak variation of lead impedance
Delta Accuracy
..........................................................
±
10 %
Baseline
..................................................................... 500, 1000, 1500, 2000
Ω
, Leads I, II, III
Accuracy Baseline
....................................................
±
5 %
Respiration Lead
....................................................... LA or LL
Apnea Selection
........................................................ OFF, 12, 22, or 32 seconds (one-time events), or continuous (Apnea ON = respiration OFF)
Power-On Default
...................................................... 20 BrPM, delta 1.0
Ω
, 1000-
Ω
baseline
Blood Pressure
Input/output Impedance
........................................... 300
Ω
±
10 %
Exciter Input Range
.................................................. 2.0 V to 16.0 V rms