ProSim™ 2/3
Users Manual
24
Other Functions
The Product also simulates respiration and temperature.
This section contains the procedures to set up the Product
for these two functions.
Respiration Functions
The Respiration function lets you set five parameters of
the respiratory waveform: Rate, Impedance, Baseline
Impedance, Lead selection (left arm or left leg), and
Apnea. To set the respiration waveform:
1. In
the
Home
screen, push the navigation buttons to
move the highlight to
RESP
.
2. Push
to show the respiration screen in the
display.
See the Function Navigation and Parameter Selection
section to learn how to set the respiration parameter
values. See the detailed specifications for range of
parameter values. These parameters change the
waveform that appears at the ECG posts along the top of
the Product.
When the Apnea parameter is set to 12, 22, or 32, the
apnea event starts immediately. When the event ends, the
parameter is set to Off. You must set the parameter to 12,
22, or 32 to start another apnea event.
The values set for the baseline and lead parameters when
the Product is turned off, become the power-up default
values.
Temperature
Temperatures simulated by the Product are compatible
with Yellow Springs, Inc. (YSI) Series 400 and 700 probes.
The type of cable connected to the temperature jack sets
the type of temperature probe simulated. Connect the
temperature input of the UUT to the Temperature jack as
shown in Figure 14.
To set temperature:
1. In
the
Home
screen, push the navigation buttons to
move the highlight to
TEMP
.
2. Push
to show the temperature screen in the
display.
See the Function Navigation and Parameter Selection
section to learn how to set the temperature parameter
value. See the detailed specifications for range of
parameter values. These parameters change the
temperature signal at the temperature connector.