ProSim™ 2/3
Users Manual
46
Exciter-Input Frequency Range
.............................. DC to 5000 Hz
Transducer Sensitivity
............................................ 5
μ
V/V/mmHg or 40
μ
V/V/mmHg
Pressure Accuracy
..................................................
±
(2 % of s 2 mmHg) (Valid for dc excitation only)
Static Levels, Channel 1
.......................................... -10, 0, 80, 160, 240, 320, 400 mmHg
Static Levels, Channel 2
.......................................... -10, 0, 50, 100, 150, 200, 240 mmHg
Static Levels, Channel 3 (ProSim 3 only)
.............. -5, 0, 20, 40, 60, 80, 100 mmHg
Static Levels, Channel 4 (ProSim 3 only)
.............. -5, 0, 20, 40, 60, 80, 100 mmHg
Dynamic Waveforms, Channel 1
............................ Arterial: 120/80
Radial artery: 120/80
Left ventricle: 120/00
Right ventricle: 25/00
Dynamic Waveforms, Channel 2
............................ Arterial: 120/80
Radial artery: 120/80
Left ventricle: 120/00
Right atrium (central venous or CVP): 15/10
Right ventricle: 25/00
Pulmonary artery: 25/10
Pulmonary-artery wedge: 10/2
Left atrium: 14/4
Dynamic Waveforms, Channel 3 (ProSim 3 only)
. Arterial: 120/80
Radial artery: 120/80
Left ventricle: 120/00
Right atrium (central venous or CVP): 15/10
Right ventricle: 25/00
Pulmonary artery: 25/10
Pulmonary-artery wedge: 10/2
Left atrium: 14/4
Dynamic Waveforms, Channel 4 (ProSim 3 only)
. Swan-Ganz sequence:
Right atrium (CVP)
Right ventricle RV)
Pulmonary artery (PA)
Pulmonary-artery wedge (PAW)
Respiration Artifact
................................................. BP delta changes from 3 mmHg to 16 mmHg
Output Connector
.................................................... DIN 5-Pin
Power-On Default
..................................................... 0 mmHg