Vital Signs Simulator
Remote Operation
27
When connected to a PC with a Windows operating
system, the Product will communicate through a PC COM
port. Make sure the COM port parameters are set to:
•
9600
Baud
•
No
Parity
•
8 data bits
•
1 stop bit
•
Hardware handshake set to off
Remote Commands
A remote command is made up of alphanumeric
characters. The first character of a command must be
alphabetic. The alphabetic characters can be upper or
lower case.
•
Special
characters
are:
•
Carriage return (CR)
•
Line feed (LF)
•
Space
(SP)
•
Backspace
(BS)
•
Escape
(ESC)
The Product will do a command when it receives a
carriage return and/or line feed. Alphabetic characters are
not case sensitive. When you type in a command, the
backspace deletes the last recorded character and the
escape key discards the complete command. When a
command is complete, the Product sends a response that
ends with a carriage return and line feed to the PC. Unless
other data is sent back from the Product, the response is
“OK” if the command is accepted by the Product. When a
command is not accepted by the Product, an error code
shown in Table 8 is sent to the PC.
Table 8. Error Codes
Error Code
Description
ERR=00
No commands allowed at this time
ERR=01 Unknown
command
ERR=02 Illegal
command
ERR=03 Illegal
parameter
ERR=04 Data
corrupted
ERR=05 Unknown
error
ERR=06
Option not installed
ERR=07 Incorrect
password
While the Product is operated from the front panel (local
mode) the remote interface will not respond to a command
until the command REMOTE is sent to the Product through
the USB port.