background image

 

HF132PEN_v0501_05

 

Page  18

4. INSTALLATION 

 

4.1. GENERAL 

DIRECTIONS 

 

  Mounting, electrical installation, start-up and maintenance of this instrument may only be carried 

out by trained personnel authorized by the operator of the facility. Personnel must read and 
understand this Operating Manual before carrying out its instructions. 

  The F132-P may only be operated by personnel who are authorized and trained by the operator 

of the facility. All instructions in this manual are to be observed. 

  Ensure that the measuring system is correctly wired up according to the wiring diagrams. 

Protection against accidental contact is no longer assured when the housing cover is removed 
or the panel cabinet has been opened (danger from electrical shock). The housing may only be 
opened by trained personnel. 

  Take careful notice of the " Safety rules, instructions and precautionary measures " at the front 

of this manual. 

 
 

4.2. INSTALLATION 

SURROUNDING CONDITIONS 

 

 
Take the relevant IP classification of the casing into account (see manufactures plate). Even an IP67 
(NEMA 4X) casing should NEVER be exposed to strongly varying (weather) conditions.  
When panel-mounted, the unit is IP65 (NEMA 4X)! 
When used in very cold surroundings or varying climatic conditions, take the necessary precautions 
against moisture by placing a dry sachet of silica gel, for example, inside the instrument case. 
 

 
 
Mount the F132-P on a solid structure to avoid vibrations. 
 
 

Содержание F132-P

Страница 1: ...Namur and coil Status inputs start and stop Control output two control outputs for one and two stage control Communication Printer interface for receipt printing Options Intrinsically Safe F Series Fi...

Страница 2: ...of according to local regulations regarding waste electronic equipment If a battery is present in this product it should be disposed of separately The separate collection and recycling of your waste e...

Страница 3: ...ocedures which if not performed correctly may lead to personal injury a safety hazard or damage of the F132 P or connected instruments A caution indicates actions or procedures which if not performed...

Страница 4: ...15 4 Display 15 5 Power management 15 6 Flowmeter 16 7 Relay output 16 8 Print settings 16 9 Others 17 4 Installation 18 4 1 General directions 18 4 2 Installation surrounding conditions 18 4 3 Dimens...

Страница 5: ...o process all types of flowmeter signals receipt printer interface to print completed batch receipts Flowmeter input This manual describes the unit with a pulse type input from the flowmeter P version...

Страница 6: ...nute Receipt information After each completed batch a receipt is printed on an external device Additional receipts can be printed on demand Below example shows a printed receipt The arrows indicate th...

Страница 7: ...ot be continued The arrow key is used to display the total accumulated total and batch counter as well as a menu option to print a batch receipt After PROG has been pressed the arrow key is used to se...

Страница 8: ...once the batch process will be temporarily interrupted the actual values are not lost On the display the PAUSE indicator will flash From this stage the batch process can be resumed with the START key...

Страница 9: ...is started READY goes off and RUN is displayed During this time you cannot print receipts 2 4 OPERATOR ALARMS Low battery alarm When the battery voltage drops it must be replaced At first the low bat...

Страница 10: ...ved Ensure that the measuring system is correctly wired up according to the wiring diagrams The housing may only be opened by trained personnel Take careful notice of the Safety rules instructions and...

Страница 11: ...a unique number which is displayed below SETUP at the bottom of the display The number is a combination of two figures The first figure indicates the function group and the second figure the sub func...

Страница 12: ...a is altered but ENTER is not pressed then the alteration can still be cancelled by waiting for 20 seconds or by pressing ENTER for three seconds the PROG procedure will be left automatically and the...

Страница 13: ...LCD UPDATE fast 1 sec 3 sec 15 sec 30 sec off 52 BATTERY MODE operational shelf 6 FLOWMETER 61 SIGNAL npn npn_lp reed reed_lp pnp pnp_lp namur coil_hi coil_lo act_8 1 act_12 act_24 7 RELAYS 71 RELAYS...

Страница 14: ...he K factor the more accurate the functioning of the system will be Example 1 Calculating the K factor Let us assume that the flowmeter generates 2 4813 pulses per liter and the selected unit is cubic...

Страница 15: ...n be enabled disabled NO FLOW TIME 32 In case of a failing signal this function determines the period of time after which an alarm has to be triggered The minimum monitoring time is 0 1 second maximum...

Страница 16: ...e input 12 VDC 4K 10KHz External power required ACT_24 Active pulse input 24 VDC 3K 10KHz External power required 7 RELAY OUTPUT Two control outputs are available to control relays or valves Relay 2 c...

Страница 17: ...CR causes a carriage return but no line feed LF causes a line feed but no carriage return Choose CR LF to get both default STARTROW 84 Enter a number of empty lines to advance before printing the rece...

Страница 18: ...contact is no longer assured when the housing cover is removed or the panel cabinet has been opened danger from electrical shock The housing may only be opened by trained personnel Take careful notice...

Страница 19: ...mm 30mm 22 5mm M20 x 1 5 M16 x 1 5 M16 x 1 5 30mm 30mm 22 50mm M20 x 1 5 22 5mm M20 x 1 5 M20 x 1 5 25mm 25mm 22 5mm 1 2 NPT 0 9 3x 1 2 NPT 0 12 0 12 0 9 HA HM HN HO HP HT HU 6 x M12 12mm 12mm 24mm 24...

Страница 20: ...box flat bottom HE HF HG HH D 12mm 12mm 12mm 24mm 24mm 36mm 36mm 14mm 17mm 22 5mm 30mm 30mm D 16mm D 20mm 0 9 D 22mm 0 866 22 5mm 25mm 25mm D 20mm D 20mm D 16mm HC 75 mm 2 95 118 mm 4 65 104 mm 4 09...

Страница 21: ...e to impact or friction sparks between the enclosure and iron steel is excluded Do ground the aluminum enclosure properly as indicated if the F132 P has been supplied with the 115 230V AC power supply...

Страница 22: ...V The voltage is selected with the three switches inside the enclosure Warning be sure that all the leads to the terminals are disconnected from the unit when the internal plastic protection cover has...

Страница 23: ...ard configuration F132 P and options REMARKS TERMINAL CONNECTORS Terminal GND 01 02 Power Supply only available with type PD PF or PM TYPE SENSOR SUPPLY Terminal backlight T YPE AA T YPE AU Type OA Ty...

Страница 24: ...active 24V DC signal according to the functions R1 and R2 is available with this option Max driving capacity 50mA 24V per output Requires power supply type PD PF PM Type OR A mechanical relay output...

Страница 25: ...ter signals can be connected to the unit pulse active pulse or sine wave coil The screen of the signal wire must be connected to the common ground terminal 09 unless earthed at the sensor itself The m...

Страница 26: ...s the maximum input frequency read par 3 2 3 A sensor supply voltage of 8 1 12 or 24V DC can be provided with power supply type PD PF PM For a signal detection level of 50 of the supply voltage please...

Страница 27: ...powered F132 P is unable to power the Namur sensor However a 8 2V sensor supply voltage terminal 11 can be provided with power supply type PD PF PM Terminal 12 13 external START With this function the...

Страница 28: ...l printer interface based on RS232 length of cable max 15 meters or RS485 length of cable max 1200 meters Fig 13 Overview terminal connectors printer interface When using the RS232 communication optio...

Страница 29: ...ns and precautionary measures in the front of this manual Safety Instructions When two or more active intrinsically safe circuits are connected to the indicator in order to prevent voltage and or curr...

Страница 30: ...r supplied by the internal supply option PC or the external supply connected to terminals 0 and 1 option PD or the circuit supply connected to terminals 7 and 8 option AP The maximum values for any of...

Страница 31: ...s the same voltage as connected to terminal 01 Terminal 11 this terminal offers a 3 2V or 8 1V to power the sensor This voltage is selected with the switch es inside the enclosure To access the voltag...

Страница 32: ...type PX 8 30 V DC 7 8 Common ground Main supply Control output type OT passive transistor e g relay e g relay Note above values are safety values Consult the technical specification for operational v...

Страница 33: ...cated Ex label are certified for replacement and use in hazardous areas Batteries for use in safe areas have no Ex label DO NOT EXCHANGE Using the wrong type of battery can pose a SERIOUS RISK For use...

Страница 34: ...ernal atmosphere of the F132 P in such a way that no condensation will occur for example by placing dry silica gel sachet in the casing just before closing it Furthermore it is required to replace or...

Страница 35: ...s Type HD Type HE Type HF Type HG Type HJ Type HH Type HK ABS enclosure Type HS Dimensions 130 x 120 x 75mm 5 10 x 4 72 x 2 95 LxHxD IP67 NEMA4X Drilling 2x PG9 1x M20 Drilling 2x M16 1x M20 Drilling...

Страница 36: ...EX II 2 GD EEx d IIB T5 Weight appr 15kg Dimensions of enclosure 350 x 250 x 200mm 13 7 x 9 9 x 7 9 LxHxD Environment Electromagnetic compatibility Compliant ref EN 61326 1997 EN 61010 1 1993 Flowmet...

Страница 37: ...d Addressing maximum 255 addresses Type CB RS232 Type CH RS485 2 wire Type CI RS485 4 wire Type CT TTL Intrinsically Safe communication Type CX no communication Operator functions Functions enter a ba...

Страница 38: ...ier Printer does not print If no receipt is printed upon request Reprint yes no or when a batch has completed check power and paper supply of printer printer self test to assure its operable see print...

Страница 39: ...25 power supply intrinsically safe 31 preset max batch size 14 Preset batch maximum 8 enter batch value 7 printer 6 9 13 16 38 Problem solving 38 Pulse signal active pulse 26 Pulse signal NPN PNP 26...

Страница 40: ...DWELL F132 P 83 lineterm CR LF 84 startrow 2 85 endrow 2 86 formfeed disable 87 dateformat Y M D 9 OTHERS 91 model F132 P F132 P F132 P 92 software version 02 05 ____ 02 05 ____ 02 05 ____ 93 serial n...

Отзывы: