Patient Therapy Controller (PTC
®
) Guide, For use with Prometra
®
Programmable Infusion Systems
26
Appendix
Technical Specifications and EMC Tables (continued)
Table 3 – Guidance and Manufacturer’s declaration – Electromagnetic Immunity –
for Patient Therapy Controller
The Patient Therapy Controller is intended for use in the electromagnetic
environment specified below. The customer or the user of this equipment should
assure that it is used in such an environment.
Immunity
Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment -
Guidance
Conducted
RF
IEC 61000-
4-6
Radiated RF
IEC 61000-
4-3
3 V
0.15 MHz – 80
MHz
6 V in ISM
bands and
amateur radio
bands between
0.15 MHz and
80 MHz
80 % AM at 1
kHz
10 V/m
80 MHz – 2.7
GHz
80 % AM at 1
kHz
3 V
0.15 MHz –
80 MHz
6 V in ISM
bands and
amateur
radio bands
between
0.15 MHz
and 80 MHz
80 % AM at
1 kHz
10 V/m
80 MHz –
2.7 GHz
80 % AM at
1 kHz
Portable and mobile RF
communications equipment
should be used no closer to any
part of this equipment, including
cables, should be used no closer
than 30 cm (12 inches) to any
part of the Patient Therapy
Controller or Clinician
Programmer. Otherwise,
degradation of the performance
could result
Interference may occur in the
vicinity of equipment marked with
the following symbol: