Patient Therapy Controller (PTC
®
) Guide, For use with Prometra
®
Programmable Infusion Systems
21
be recharged.
• You cannot replace the rechargea
ble battery inside the device.
• If the battery is no longer charging, then contact your physician.
• Over time, if you do not use your Patient Therapy Controller, the battery will deplete and must be recharged before use.
•
The life of the Patient Therapy Controller’s battery will vary depending on use. If there is a suspected issue with the battery, contact your
physician to coordinate getting a replacement.
Service and Maintenance (continued)
Service
The Patient Therapy Controller has no serviceable parts. For technical issues or questions about the programmer, please contact your physician or
contact Flowonix Medical‘s Technical Solutions Department.
Handling and Care
•
Do not drop the Patient Therapy Controller or subject it to severe shock. If the device becomes damaged, contact your physician.
• Keep the device dry; humidity and all types of liquids may cause damage.
•
Do not turn on the device if it is wet. If the device is already on, turn it off (if the device will not turn off, leave it as-is). Then, dry the device with a
towel and contact your physician.
Caution: If device becomes wet, do not use as there is a possible electrical shock hazard.
•
Do not store the device in hot or cold areas. Only use the Patient Therapy Controller between
68° F and 104° F
, The device can explode if left
inside a closed vehicle, as the inside temperature can reach up to 176° F.
• Do not expose the device to direct sunlight for extende
d periods of time (such as on the dashboard of a car).
• Storage temperature should be between 32° F and 104° F.
•
Do not store the device near magnetic fields. The device may malfunction or the battery may discharge from exposure to magnetic fields.
•
Do not use carrying cases or accessories with magnetic closures or allow the device to encounter magnetic fields for extended period of time,
including exposure to MRI machines.
If the Patient Therapy Controller becomes dirty:
• Wipe your device or charger w
ith a towel or cloth. Do not spray any liquids on the device or allow droplets to enter the charging port.
• Do not use chemicals or detergents.
• Do not allow excessive moisture to enter the device. Do not immerse the device in any type of liquid.
Sterilization
Do not sterilize the Patient Therapy Controller. If a sterile barrier is desired, place the device in a clear sterile bag or sleeve.