Patient Therapy Controller (PTC
®
) Guide, For use with Prometra
®
Programmable Infusion Systems
20
Battery charging. PTC
disabled. Unplug to
use.
The Patient Therapy
Controller is charging. Your
Patient Therapy Controller
cannot be used while
charging.
Allow recharging to complete on the
device. Unplug your Patient Therapy
Controller from the charger in order
to use.
Troubleshooting
PTC® Pump Error Messages
If the Patient Therapy Controller receives an error message from the Programmable Pump, the Patient Therapy Controller will display the pump
error message before displaying the standard programming screens. Each error message will be associated with a unique numeric code displayed
along with the error message. The numeric code helps the Technical Solutions Department identify the issue with the pump. Below is an example
of what an error message would look like. In all cases, contact your physician or Technical Solutions Department for further assistance.
Service and Maintenance
The following applies to the Patient Therapy Controller.
Warning: All batteries should be disposed of according to local and federal regulation.
•
If you are no longer going to use the Patient Therapy Controller, you should return it to your physician. Your physician will arrange to have it to
returned to Flowonix Medical.
Batteries for PTC® (Rechargeable/Internal)
• Use the supplied Micro USB Charger (AC adapter) to charge the internal battery for the touchscreen using a UL approved powe
r source.
• The charging USB port is located at the bottom of the device.
•
When the battery reaches approximately 10% of its charge, a message will appear on the screen indicating that the battery is low and needs to