Appendix iii
Programming Error Messages
Note: All Error messages are followed by the phrase:
"Push ENT to continue".
These messages occur when attempting to go to the run-
ning mode without entering necessary information or
entering a value that is not allowed.
Messages
Temp hi-lo lim err:
Temperature input parameters have not
been entered or a low limit is higher than a high limit.
Press hi-lo lim err:
Pressure input parameters have not been
entered or a low limit is higher than a high limit.
Flow hi-lo lim err:
Flow input parameters have not been en-
tered or a low limit is higher than a high limit.
Need more flow info:
All of the Flow input parameters have
not been entered.
Z factor error:
Z factor has been set to zero, change to any
number greater than zero.
Scale factor error:
K factor has been set to zero, change to any
number greater than zero.
No value stored yet:
The unit still has the null value in it from
initialization. Enter a value.
Bad lock code:
Attempt has been made to enter 00000 as the
lockout code, use any other number than 00000 to lock out computer.
Syntax error:
Time or date was not entered in proper format.
Refer to Section 3.12 for proper format.
Bad Time value:
Invalid time entered. Refer to Section 3.12 for
proper format.
Bad Date value:
Invalid date entered. Refer to Section 3.12 for
proper format.
Bad ID value:
Invalid ID entered. Refer to Section 3.13 for
proper format.