Advanced Adjustments (continued)
Step Eight
UPDATE AN EXISTING FALSE ECHO CURVE
This function enables an Existing False Echo Curve to be updated under the circumstances that the curve was
created when the level was higher than an empty tank condition.
Note: This function should only be performed
to update an Existing False Echo Curve when the level is BELOW the original False Echo Curve. Do not use
this function to update an Existing False Echo Curve when the level is above the original False Echo Curve.
Before starting, measure and note the exact distance from the sensor’s measurement location to the media
surface. Setting the distance value too large or too short can force the sensor into ignoring the true level.
Original False Echo
Curve
1) From the Main Screen, press
Enter
to advance into
the Main Menu.
2)
Press
Right Arrow
repeatedly until the arrow is next
to Service.
3)
Press
Enter
to advance into Service Menu and Echo
Curve will appear.
4)
Press
Enter
to make a change to the Echo Curve
settings.
5) Press
Right Arrow
repeatedly until the arrow is next to
Update.
6)
Press
Enter
to advance into Update. Enter the actual
distance from the sensor’s measurement location to
the media surface.
7)
Use the
Right Arrow
to move one segment to the
right
.
The
Right Arrow
will scroll left to right and then
back to the first segment.
8)
Use the
Up Arrow
to increase the value of the number
highlighted. The
Up Arrow
will scroll from 0 to 9 and
back again.
9) When the value is correct, press
Enter
to save the
setting and begin the False Echo Curve mapping.
The process may take a few minutes. When
complete, the display will return to the Echo Curve
screen.
10) When done,
press
ESC
to return to the Service Menu
and press
ESC
a second time to return to the Main
Screen.
11)
From the Main Screen, press
ESC
and the Echo
Curve Screen will appear. Follow the Check Echo
Curve procedure described at the end of Section Five
to confirm that the sensor is performing correctly.
Good Update
Bad Update