23
OPERATION (continued)
USER CONFIGURATION MODE (Advanced Options - continued)
Output K-Factor Menu
(Requires Pulse Access or 4-20mA Daughter Board Module)
The scaled pulse output K-factor menu allows the user to enter a K-factor for the
pulse output of the meter or advance focus to the next menu.
The software will read the pulse output K-factor from the computer and display it on
the top row. The left most digit will be in focus to indicate that it can be changed. The
decimal can also be moved (see Figure 28).
When advancing focus to the next menu
the software will save the Scaled Pulse K-
factor to the computer.
Figure 28
4-20mA Menu (Requires 4-20mA Daughter Board Module)
Zero Setting
This figure will flash for 1
second before allowing the user to change
the 4-20mA zero setting (see Figure 29-1).
This setting will already be configured for the
minimum flowrate of the meter this Q9 was
attached to when shipped from the factory.
Figure 29-1
The button presses required to change this
menu item are the same as in the Pipe
Diameter Menu above (see Figure 29-2)
Figure 29-2
Span Setting
This figure will flash for 1
second before allowing the user to change the
4-20mA span setting (see Figure 29-3). This
setting will already be configured for the
maximum flowrate of the meter this Q9 was
attached to when shipped from the factory.
Figure 29-3
24
TROUBLESHOOTING