Action
Step
Measure the radiation intensity (= apparent temperature) from the reflecting source
using the following settings:
■
Emissivity: 1.0
■
D
obj
: 0
You can measure the radiation intensity using one of the following two methods:
10589003;a2
Figure 16.3 1 = Reflection source
3
➲
Please note the following:
Using a thermocouple to measure reflecting temperature is not recommended for
two important reasons:
■
A thermocouple does not measure radiation intensity
■
A thermocouple requires a very good thermal contact to the surface, usually by
gluing and covering the sensor by a thermal isolator.
16.2.1.1.2
Method 2: Reflector method
Action
Step
Crumble up a large piece of aluminum foil.
1
Uncrumble the aluminum foil and attach it to a piece of cardboard of the same
size.
2
Put the piece of cardboard in front of the object you want to measure. Make sure
that the side with aluminum foil points to the camera.
3
Set the emissivity to 1.0.
4
16
Publ. No. 1557882 Rev. a156 – ENGLISH (EN) – February 28, 2006
133
16 – Thermographic measurement techniques
Содержание ThermaCAM B2
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 7: ...ThermaCAM B2 User s manual Publ No 1557882 Rev a156 ENGLISH EN February 28 2006...
Страница 10: ...Reg No Status Designation 29 233 400 Pending U S x Publ No 1557882 Rev a156 ENGLISH EN February 28 2006...
Страница 16: ...INTENTIONALLY LEFT BLANK 1 2 Publ No 1557882 Rev a156 ENGLISH EN February 28 2006 1 Warnings cautions...
Страница 26: ...INTENTIONALLY LEFT BLANK 4 12 Publ No 1557882 Rev a156 ENGLISH EN February 28 2006 4 Packing list...
Страница 28: ...INTENTIONALLY LEFT BLANK 5 14 Publ No 1557882 Rev a156 ENGLISH EN February 28 2006 5 System overview...
Страница 110: ...INTENTIONALLY LEFT BLANK 12 96 Publ No 1557882 Rev a156 ENGLISH EN February 28 2006 12 Maintenance cleaning...
Страница 164: ...INTENTIONALLY LEFT BLANK 18 150 Publ No 1557882 Rev a156 ENGLISH EN February 28 2006 18 Theory of thermography...
Страница 191: ......