FLIR DM166 USER MANUAL Document Identifier: DM166-en-US_AA
33
AC Current*
RANGE and RESOLUTION
Accuracy
Burden Voltage
45H
Z
~
440H
Z
600.0
A
1) 2)
, 6000
A
2)
1.5% + 3d
0.1mV/
A
60.00mA
1) 2)
, 600.0 mA
2) 3)
1.0% + 3d
1.9mV/mA
6.000A
1) 4) 5)
, 10.00A
4) 5) 6)
0.04V/A
1) Non-zero residual readings
5d may appear as a zero ampere input when battery indication is
25%
2)
A/mA DC accuracy is affected by extreme interior temperature of the meter. For rated accuracies, allow
linear proportional cool down intervals from 6 to 20 minutes after measuring currents from 3 to 10A
continuously via A-input.
3) <400mA continuous; >400mA for <20 minutes ON per >5 minutes OFF
4) 10A continuous up to ambient 35oC; <15 minutes ON per >5 minutes OFF @ 95~122°F (35°C ~ 50°C)
5) >10A to 20A for <30 seconds ON per >5 minutes OFF
6) Unspecified @ <0.5A
Temperature
RANGE and RESOLUTION
Accuracy
1) 2) 3)
-40.0
o
C ~ 0.0
o
C
1% + 1.5
o
C
0.0
o
C ~ 100.0
o
C
1% + 1
o
C
100.0
o
C ~ 400.0
o
C
-40.0
o
F ~ 32.0
o
F
1% + 3
o
F
32.0
o
F ~212.0
o
F
1% + 2
o
F
212.0
o
F ~ 752.0
o
F
1) Accuracies assume meter interior has the same temperature as the ambient (isothermal stage) for correct
junction voltage compensation. Allow enough time to reach the isothermal stage for a significant change of
ambient temperature. It can take up to an hour for changes > 9°F (5°C).
2) Type-K thermocouple range & accuracy not included
3) For isothermal stage, allow linear proportional cool down intervals from 9 to 30 minutes after measuring
currents from 3 to 10A continuously via A-input.
Line Frequency
Function
Sensitivity (Sine RMS)
Range
60mV, 600mV
50mV
10Hz - 50kHz
6V
5V
10Hz - 50kHz
60V
10V
10Hz - 50kHz
600V
50V
10Hz - 1kHz
600V VFD
50V
10Hz-400Hz
600
A, 6000
A
500
A
10Hz - 5kHz
60mA, 600mA
50mA
10Hz - 5kHz
6A, 10A
8A
50Hz - 1kHz
Accuracy: 0.03% + 2d