A Menu Structure
92
UMFLUXUS_F5WDV1-1EN, 2016-04-08
input of the offset
range: 0...900 m/s
0: the default value 300 m/s is used
This display will only be indicated if the SuperUser mode is activated.
x
activation of the overflow of the totalizers
x
activation of the taking-over of the totalizer values after a restart of the measure-
ment
x
input of the time interval without any valid measured values after which the
transmitter recognizes a long measurement failure
0: the default value 30 s is used
This display will only be indicated if the SuperUser mode is activated.
x
input of the number of decimal places for the totalizers:
Automatic
: dynamic adjustment
Fixed to x digit
: 0...4 decimal places
This display will only be indicated if the SuperUser mode is activated.
x
input of the time duration after which the flow totalizer display toggles
between positive and negative flow direction
x
activation of the manual reset of the totalizers
This display will only be indicated if the SuperUser mode is activated.
x
activation of the display of the sum of the totalizers
This display will only be indicated if the SuperUser mode is activated.
x
activation of the display of the last valid measured value
This display will only be indicated if the SuperUser mode is activated.
x
activation of the turbulence mode
x
SYSTEM settings\Proc. outputs
selection of the displays for the setting of the outputs of the transmitter
selection of the output to be installed
SYSTEM settings\Storing
(only with the optional RS485)
selection of the displays for the storing of measured values in the data logger
setting of the overflow behavior of the data logger
x
INIT-
resistant
Bad soundspeed
offset: +321 m/s
Quant. wrapping
off
>ON<
Quantity recall
off
>ON<
Do not total. if
no meas.>
0 s
Total digits
↕
Automatic
Toggle totalizer
0
s
3xC clear totals
off
>ON<
Show
Q
off
>ON<
Keep display val
off
>ON<
Turbulence mode
off
>ON<
SYSTEM settings
↕
Proc. outputs
Install Output
↕
Current I1
SYSTEM settings
↕
Storing
Ringbuffer
off
>ON<
Содержание FLUXUS WD100
Страница 1: ...User Manual UMFLUXUS_F5WDV1 1EN Ultrasonic Flowmeter for Water FLUXUS WD100 FLUXUS WD200...
Страница 4: ...FLUXUS WD100 WD200 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 8: ...FLUXUS WD100 WD200 Table of Contents 8 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 103: ...D Declarations of Conformity FLUXUS WD100 WD200 UMFLUXUS_F5WDV1 1EN 2016 04 08 103 D Declarations of Conformity...
Страница 104: ...FLUXUS WD100 WD200 D Declarations of Conformity 104 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 105: ......
Страница 106: ......