15 Outputs
UMFLUXUS_F5WDV1-1EN, 2016-04-08
77
15.1.3
Function Test
The function of the installed output can now be tested. Connect a multimeter with the installed output.
Test of the analog output
Test of the binary outputs
15.2
Error Value Delay
The error value delay is the time interval after which the error value will be transmitted to the output in case no valid mea-
sured values are available. The error value delay can be entered in the program branch
Output Options
if this menu
item has previously been activated in the program branch
Special Funct.
. If the error value delay is not entered, the
damping factor will be used.
15.3
Activation of an Analog Output
The current output is tested in the display. Enter a test value. It has to be within the output
range. Press ENTER.
If the multimeter displays the entered value, the output functions correctly.
Press
yes
to repeat the test,
no
to return to the
SYSTEM settings
. Press ENTER.
Select
Opto-Relay OFF
or
Open collect OFF
in the scroll list
Output Test
to test
the de-energized state of the output. Press ENTER. Measure the resistance at the output.
The value has to be high ohmic.
Select
yes
. Press ENTER.
Select
Opto-Relay ON
or
Open collect. ON
in the scroll list
Output Test
to test the
energized state of the output. Press ENTER. Measure the resistance at the output. The val-
ue has to be low ohmic.
Select
yes
to repeat the test,
no
to return to
SYSTEM settings
. Press ENTER.
Select
Special Funct.\SYSTEM settings\Dialogs/Menus\Error-val. de-
lay.
Select
Damping
if the damping factor is to be used as the error value delay. Select
Edit
to
activate the menu item
Error-val. delay
in the program branch
Output Options
.
From now on, the error value delay can be entered in the program branch
Output Op-
tions
.
Note!
An output can only be activated in the program branch
Output Options
if it has previously been in-
stalled.
Select the program branch
Output Options
. Press ENTER until
Current Loop
is dis-
played.
Select
yes
to activate the output. Press ENTER.
I1:Output Test
4
mA
I1= 4.0 mA
Again? no
>YES<
B1:Output Test
↕
Opto-Relay OFF
B1=OFF
AGAIN? no
>YES<
B1:Output Test
↕
Opto-Relay ON
B1=ON
AGAIN? no
>YES<
Error-val. delay
>DAMPING<
edit
Error-val. delay
10
s
Current Loop
I1: no
>YES<
Содержание FLUXUS WD100
Страница 1: ...User Manual UMFLUXUS_F5WDV1 1EN Ultrasonic Flowmeter for Water FLUXUS WD100 FLUXUS WD200...
Страница 4: ...FLUXUS WD100 WD200 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 8: ...FLUXUS WD100 WD200 Table of Contents 8 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 103: ...D Declarations of Conformity FLUXUS WD100 WD200 UMFLUXUS_F5WDV1 1EN 2016 04 08 103 D Declarations of Conformity...
Страница 104: ...FLUXUS WD100 WD200 D Declarations of Conformity 104 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 105: ......
Страница 106: ......