15 Outputs
UMFLUXUS_F5WDV1-1EN, 2016-04-08
81
If the switching condition
QUANT.
is selected in the scroll list
func
, the limit of the output have to be defined:
A positive limit will be compared to the totalizer value for the positive flow direction.
A negative limit will be compared to the totalizer value for the negative flow direction.
The comparison will also take place if the totalizer of the other flow direction is displayed.
15.5.3
Defining the Hysteresis
A hysteresis can be defined for the alarm output R1 to prevent a constant triggering of the alarm due to small fluctuations
of the measured values around the limit.
The hysteresis is a symmetrical range around the limit. The alarm will be activated if the measured values exceed the up-
per limit and deactivated if the measured values fall below the lower limit.
15.6
Behavior of the Alarm Outputs
15.6.1
Apparent Switching Delay
Measured values and totalizer values will be displayed rounded to two decimal places. The limits, however, will be com-
pared to the non-rounded measured values. This might cause an apparent switching delay when the measured value
changes marginally (less than two decimal places). In this case, the switching accuracy of the output is greater than the
accuracy of the display.
switching condition:
QUANT.
Enter the limit of the totalizer. Press ENTER.
The alarm will switch if the measured value reaches the limit.
Note!
The unit of measurement of the limit corresponds to the unit of measurement of the selected physical
quantity.
If the unit of measurement of the physical quantity is changed, the limit has to be converted and en-
tered again.
Example 1:
physical quantity: volumetric flow rate in m³/h
Quantity Limit:
1 m³
Example 2:
physical quantity: volumetric flow rate in m³/h
Low Limit:
60 m³/h
The unit of measurement of the physical quantity is changed to m³/min. The new limit to be entered is
1 m³/min.
Example:
High Limit:
30 m³/h
Hysterese
: 1 m³/h
The alarm will be triggered at values > 30.5 m³/h and deactivated at values < 29.5 m³/h.
switching condition:
MIN or MAX
Enter the value for
Hysterese
.
or
Enter 0 (zero) to work without a hysteresis.
Press ENTER.
Quantity Limit:
1.00
m3
R1 Hysterese:
1.00
m3/h
Содержание FLUXUS WD100
Страница 1: ...User Manual UMFLUXUS_F5WDV1 1EN Ultrasonic Flowmeter for Water FLUXUS WD100 FLUXUS WD200...
Страница 4: ...FLUXUS WD100 WD200 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 8: ...FLUXUS WD100 WD200 Table of Contents 8 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 103: ...D Declarations of Conformity FLUXUS WD100 WD200 UMFLUXUS_F5WDV1 1EN 2016 04 08 103 D Declarations of Conformity...
Страница 104: ...FLUXUS WD100 WD200 D Declarations of Conformity 104 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 105: ......
Страница 106: ......