5 Selection of the Measuring Point
UMFLUXUS_F5WDV1-1EN, 2016-04-08
21
5.3
Selection of the Measurement Arrangement Taking into Account the Measuring
Range and the Measuring Conditions
disturbance source:pump
supply line: l
≥
50 D
Diagonal arrangement with 1 beam
Reflection arrangement with 1 beam
• wider flow velocity and sound speed range compared to
the reflection arrangement
• use in the presence of deposits on the inner pipe wall or
with strongly attenuating fluids (only 1 sound path)
• smaller flow velocity and sound speed range compared to
the diagonal arrangement
• transverse flow effects are compensated for because the
beam crosses the pipe in 2 directions
• higher accuracy of measurement because the accuracy
increases with the number of sound paths
Tab. 5.2: Recommended distance from disturbance sources; D – nominal pipe diameter at the measuring point, l – recommended distance
l