5 Selection of the Measuring Point
UMFLUXUS_F5WDV1-1EN, 2016-04-08
19
5.2
Undisturbed Flow Profile
Some flow elements (elbows, slide valves, valves, control valves, pumps, reducers, diffusers, etc.) distort the flow profile
in their vicinity. The axisymmetrical flow profile needed for correct measurement is no longer given. A careful selection of
the measuring point helps to reduce the impact of disturbance sources.
It is most important that the measuring point is chosen at a sufficient distance from any disturbance sources. Only then it
can be assumed that the flow profile in the pipe is fully developed. However, measuring results can be obtained even if the
recommended distance to disturbance sources cannot be observed for practical reasons.
Recommended straight inlet and outlet pipe lengths for different types of flow disturbance sources are shown in the ex-
amples in Tab. 5.2.
Tab. 5.2: Recommended distance from disturbance sources; D – nominal pipe diameter at the measuring point, l – recommended distance
disturbance source: 90° elbow
supply line: l
≥
10 D
return line: l
≥
5 D
disturbance source: 2x 90° elbows in same plane
supply line: l
≥
25 D
return line: l
≥
5 D
disturbance source: 2x 90° elbows in different planes
supply line: l
≥
40 D
return line: l
≥
5 D
l
l
l
l
l
l
Содержание FLUXUS WD100
Страница 1: ...User Manual UMFLUXUS_F5WDV1 1EN Ultrasonic Flowmeter for Water FLUXUS WD100 FLUXUS WD200...
Страница 4: ...FLUXUS WD100 WD200 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 8: ...FLUXUS WD100 WD200 Table of Contents 8 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 103: ...D Declarations of Conformity FLUXUS WD100 WD200 UMFLUXUS_F5WDV1 1EN 2016 04 08 103 D Declarations of Conformity...
Страница 104: ...FLUXUS WD100 WD200 D Declarations of Conformity 104 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 105: ......
Страница 106: ......