13 Settings
68
UMFLUXUS_F5WDV1-1EN, 2016-04-08
13.3
Measurement Settings
13.4
Setting the Contrast
It is possible to reset the display to fluid contrast. Enter HotCode
555000
.
13.5
Instrument Information
Select
Special Funct.\SYSTEM settings\Measuring
. Press ENTER.
Note!
The settings of the menu item
Measuring
will be stored at the end of the dialog. If the menu item is
quit before the end of the dialog, the settings will not be effective.
A lower limit for the flow velocity can be entered (see section 11.5).
An upper limit for the flow velocity can be entered (see section 11.4).
Enter
0
(zero) to deactivate the flow velocity check.
Select the overflow behavior of the totalizers (see section 11.3.1).
Select
on
to keep the previous totalizer values after a restart of the measurement.
Select
off
to reset the totalizers to zero after a restart of the measurement.
Note!
All changes will be stored at the end of the dialog.
Select
Special Funct.\SYSTEM settings\Miscellaneous
to set the contrast of
the display of the transmitter. Press ENTER.
The contrast of the display is adjusted with the follow ing keys:
increases the contrast
reduces the contrast
Note!
After an initialization of the transmitter, the display is displayed to fluid contrast.
Select
Special Funct.\Instrum. Inform.
to display information about the transmit-
ter. Press ENTER.
The type and the serial number of the transmitter will be displayed in the upper line.
The firmware version of the transmitter with date is displayed in the lower line.
Press ENTER.
SYSTEM settings
↕
Measuring
Cut-off Flow
absolut
>SIGN<
Cut-off Flow
factory
>USER<
Velocity limit
24.0
m/s
Quant. wrapping
off
>ON<
Quantity recall
off
>ON<
SYSTEM settings
↕
Miscellaneous
SETUP DISPLAY
←
CONTRAST
→
Special Funct.
↕
Instrum. Inform.
XXXX
-XXXXXXXX
V x.xx
dd.mm.yy
Содержание FLUXUS WD100
Страница 1: ...User Manual UMFLUXUS_F5WDV1 1EN Ultrasonic Flowmeter for Water FLUXUS WD100 FLUXUS WD200...
Страница 4: ...FLUXUS WD100 WD200 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 8: ...FLUXUS WD100 WD200 Table of Contents 8 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 103: ...D Declarations of Conformity FLUXUS WD100 WD200 UMFLUXUS_F5WDV1 1EN 2016 04 08 103 D Declarations of Conformity...
Страница 104: ...FLUXUS WD100 WD200 D Declarations of Conformity 104 UMFLUXUS_F5WDV1 1EN 2016 04 08...
Страница 105: ......
Страница 106: ......