12 Data Logger and Transmission of Data
UMFLUXUS_F7V4-6-2EN, 2017-10-01
113
12.1.3
Settings of the Data Logger
Select program branch
Special Funct.\SYSTEM settings\Storing
. It contains the following menu items:
• ringbuffer
• storage mode
• storing of the totalizers
• storing of the signal amplitude
• storing of the sound speed
• start of the storing
• acoustic signal during the storing
Ringbuffer
The setting of ringbuffer affects the storing of measured values as soon as the data memory is full:
• If the ringbuffer is activated, the available data logger will be halved. The oldest measured values will be overwritten.
Only the data logger memory that was free during the activation will be used by the ringbuffer. If more data logger mem-
ory is necessary, measured values in the data logger should previously be deleted.
• If the ringbuffer is deactivated, the storing of measured values will be stopped.
Storage mode
Storing of the totalizers
Storing of the signal amplitude
Storing of the sound speed
Storing of the diagnostic values
Select the behavior of the ringbuffer. Press ENTER.
Select the storage mode. Press ENTER.
If
sample
is selected, the displayed measured value will be used for storing and online
transmission of data.
If
average
is selected, the average of all values measured during a storage interval will be
used for storing and online transmission of data.
Note!
The storage mode does not affect the outputs.
Note!
Storage mode
=
average
The average of the physical quantity and other physical quantities assigned to the measuring chan-
nel, e.g., the measured temperature, will be calculated.
If the storage rate < 5 s (see section 12.1.2) is selected,
sample
will be used.
If no average could be calculated over the complete storage interval, the value will be marked as in-
valid. The ASCII file will contain "
???
" instead of invalid average values of the physical quantity and
"?UNDEF
" instead of invalid temperatures.
If
on
is selected and the data logger is activated, the amplitude of the measured signal will
be stored together with the measured values. Press ENTER.
If
on
is selected and the data logger is activated, the sound speed of the medium will be
stored together with the measured values. Press ENTER.
If
on
is selected and the data logger is activated, the diagnostic values will be stored to-
gether with the measured values. Press ENTER.
Ringbuffer
off
>ON<
Storage mode
>SAMPLE< average
Store Amplitude
off
>ON<
Store c-Medium
off
>ON<
Store diagnostic
off
>ON<