7 Connection
UMFLUXUS_F7V4-6-2EN, 2017-10-01
85
If transducers are replaced or added, the sensor module has to be replaced or added.
FLUXUS F704, FLUXUS F705
• Disconnect the power supply from the transmitter.
• For the connection of the AMP-Quick connector (connection system AS), see section 7.1.2.1.
• For the connection of the separate sensor module (connection system TS):
– Insert the sensor module into the lower row of terminal strip (see Fig. 7.1 or Fig. 7.2 and Fig. 7.51 or Fig. 7.52). The slots
SA1...SA4 are assigned to the transducers of measuring channel A, the slots SB1...SB4 are assigned to the transducers
of measuring channel B.
• Connect the transmitter to the power supply.
• Enter all parameters of the program branch
Stop Charging
.
• Press ENTER until the main menu is displayed again.
FLUXUS F706
• Disconnect the power supply from the transmitter.
• For the connection of the AMP-Quick connector (connection system AS), see section 7.1.2.1.
• For the connection of the separate sensor module (connection system TS):
– Insert the sensor module into the lower row of terminal strip (see Fig. 7.1 or Fig. 7.2 and Fig. 7.51 or Fig. 7.52). The
slots ROM A are assigned to the transducers of measuring channel A, the slots ROM B are assigned to the transduc-
ers of measuring channel B, the slots ROM C are assigned to the transducers of measuring channel C, the slots ROM
D are assigned to the transducers of measuring channel D.
• Connect the transmitter to the power supply.
• Enter all parameters of the program branch
Stop Charging
.
• Press ENTER until the main menu is displayed again.
Note!
The serial number of sensor module and transducer have to be identical. A wrong or incorrectly con-
nected sensor module will lead to incorrect measured values or to a measurement failure.
Fig. 7.51: Transmitter FLUXUS F704
Fig. 7.52: Transmitter FLUXUS F705
Abb. 7.53: Messumformer FLUXUS F706
AVS
AV
AGN ARS AR
BVS
BV
BGN BRS BR
SA1 SA2 SA3 SA4
SB1 SB2 SB3 SB4
T1a T1b S2 T2a
T3a T3b S4 T4a
T2b
T4b
T1A T1B S1 T2A
T3A T3B S3 T4A
T2B
T4B
A+ B- P1+ P2+
P4+ P5a P6a P7a
P3+
101 103 P1- P2-
P4- P5b P6b P7b
P3-
PE N(-) L(+)
sensor module
measuring channel A
sensor module
measuring channel B
AVS
AV
AGN ARS
X2
X_BV
X_BR
X_AV
X_AR
X1
AR
BVS
BV
BGN BRS BR
SA1 SA2 SA3 SA4
SB1 SB2 SB3 SB4
T1a T1b S2 T2a
T3a T3b S4 T4a
T2b
T4b
T1A T1B S1 T2A
T3A T3B S3 T4A
T2B
T4B
A+ B- P1+ P2+
P4+ P5a P6a P7a
P3+
101 103 P1- P2-
P4- P5b P6b P7b
P3-
PE
N(-)
L(+)
sensor module
measuring channel A
sensor module
measuring channel B
T1a
T1b
S2 T2a
T3a T3b
S4 T4a
T2b
T4b
T1A T1B
S1 T2A
T3A T3B
S3 T4A
T2B
T4B
A+
B-
P1+ P2+
P4+ P5a P6a P7a
P3+
101 103
P1-
P2-
P4- P5b P6b P7b
P8a P9a
P8b P9b
P3-
PE
N(-)
L(+)
AVS
AV
AGN ARS AR
BVS
BV
BGN BRS BR
CVS
CV
CGN CRS CR
DVS
DV
DGN DRS DR
X_AR
X_AV
KL3
KL2
KL1
ROM A
ROM B
ROM C
ROM D
X_BV
X_BR
X_CV
X_CR
X_DV
X_DR
sensor module
measuring channel A
sensor module
measuring channel C
sensor module
measuring channel D
sensor module
measuring channel B
Содержание FLUXUS F704**-A2 Series
Страница 4: ...FLUXUS F70x UMFLUXUS_F7V4 6 2EN 2017 10 01...
Страница 8: ...FLUXUS F70x Table of Contents 8 UMFLUXUS_F7V4 6 2EN 2017 10 01...
Страница 190: ...FLUXUS F70x D Declarations of Conformity 190 UMFLUXUS_F7V4 6 2EN 2017 10 01 D Declarations of Conformity...
Страница 191: ......
Страница 192: ......