Note: Refer to the following for more information on the Strip
Lockout function.
Once a canister code is accepted, the
analyzer will automatically recognize the Type of Strip and the
number of strips per canister should be tested. The analyzer will
keep track of the number of strips available for testing and the
strips remaining will be displayed under the lock symbol.
Continue testing as usual until the analyzer prompts that a
new canister code is required. When the analyzer prompts
that a new canister code is required, scan or manually enter a
canister code from a new canister to be used, or from the one
currently in use.
The analyzer will not accept canister codes that
have already been entered and will lockout the user until a new
canister code is entered. One canister code can be entered for
each Type of Strip, and will be managed independently of other
strip types by the analyzer.
Normal Operation, no Barcode reader
Sample/Strip Preparation
Allow the strip, urine specimen, and/or controls to reach room temperature at
15-30ºC (59-86ºF) prior to testing.
Note:
Only
AimStrip
®
and
Fisherbrand
®
Urinalysis Reagent Strips work
with the
AimStrip
®
Urine Analyzer and the
CTMI
CT-120 Analyzer.
Use of any other strips will cause inaccurate results.
Remove the strips from the closed canister. Use them as soon as possible.
Tightly close the canister after removing the strips.
Strip Test – Single Test Mode
Press
ENTER
for the strip prompt. Wait for the audible triple beep to immerse the
strip.
Once
START
is pressed, the screen below will show.