Autoclaving
Dis-assembling for Autoclaving
•
1- Unscrew Chuck Nut 1
•
2- Pull Chuck Nut 1 along with barrel cover, upper housing and piston all the way up.
•
3- Autoclave the two sub-assemblies at 0 121 C and 15 psi pressure for 10-15 mins.
Re-assembling after autoclaving
•
1- Push the Piston in the Glass Barrel gently and go all the way down. (Caution : Ensure that the Nozzle cap is removed)
•
2- Tighten chuck nut 1 by screwing it properly.
•
3- Dispenser is now ready for use. No Re-calibation is required after Autoclaving.However, a quick calibration check is
recommended.
Bottle Top Dispenser
Instruction Manual