A. Power switch
B. Light switch
C. 3 digits display with dot separation value
D. temperature LED
E. humidity LED
F. selection button <arrow up>
G. setting button
H. selection button <arrow down>
4. User setup
After you turn-on MiniLCD by pressing the
button, RED LED close to temperature icon start blinking and display
show the reference temperature.
Now by pressing
or
you can change the set temperature.
• T. incubazione [5 .. 75] (default 30.0)
This parameter define the temperature that the machine must be maintened during the incubation period. The range
of values you can set is between 5 and 75 °C or equivalent in F.
You can save the value by pressing the
button, display will show the humidity value.
Now the GREEN LED close to humidity icon start blinking and display show the humidity percentage to be observed
For modify the humidity parameter pressing
or
button.
• Umidità incubazione [10 .. 99] (default 40)
This parameter allows you to set the relative humidity to be maintained during the incubation period. The range of
values that can be set in this field is between 10% and 99%.
After editing the values MiniLCD automatically exits the edit menu if no buttons are pressed for 15 seconds by saving
the value shown.
5. Technical setup
By simultaneously pressing
and
you enter the technical menu. Its possible to read all parameters in sequence
by pressing
, button, the modify of the parameter occur by pressing
and
.
To confirm the modify of the value press again
.
MiniLCD will exit automatically after 15 seconds if no keys are pressed by saving the value shown.
List of parameters showed by pressing
button
• Unit [C,F] (default °C)
Setting the temperature units Celsius or Fahrenheit
• t.CAL [-3,0 .. +3,0 °C] (default 0,0 – step di 0,1 °C)
t.CAL [-5,4 .. +5,4 F] (default 0,0 – step di 0,1 F)
It allows to calibrate the temperature read by the probe, in Celsius or Fahrenheit depending on the setting assigned
at the previus point.
• H.Cal [-5 .. +5] (default 0 – step di 1%)
It allows to calibrate the humidity value read by the probe.
6. Control mode temperature
By pressing simultaneously
and
you can access a special menu for change the mode of control temperature.
• Controllo [PID,v ON-OFF] (default PID) Allow you to select the type of temperature control.
• PID = Control PID
• ON-OFF= Control ON-OFF
7. CONNECTION DIAGRAM
Descrizione dei collegamenti:
A = Power supply terminal block ( ELECTRIC SHOCK )
B = Power output humidifier terminal block ( ELECTRIC SHOCK )
C = Power output heater ( ELECTRIC SHOCK )
D = Hygrometer probe
E = Temperature probe
Warning: disconnect power supply before attempting and performing any conn.
13