2 INSTALLATION
2.A INSTRUCTIONS AND CHECKS ON DELIVERY
Considering the critical nature of the use of the GLOVEFAST ASEPTIC isolator and the need to keep it in
optimum condition, installation is very important.
GLOVEFAST ASEPTIC isolator are positioned on a pallet wrapped in an extensible film and contained in a
package of multi-layer strapped cardboard or a wooden cage.
After placing the isolator in its site of use, opened the package and removed the extensible film, check that
the equipment has not suffered any dents or scratches due to transport or improper handling of the package.
In case of any further transport, packing and storage by the user after the initial period of use (e.g.: change
of laboratory or factory), contact the technical assistance service or the distributor for more accurate and
precise instructions or for assistance by specialized technicians.
GLOVEFAST ASEPTIC isolator, with or without package, should be always located in a room sheltered
from rain.
2.B INSTALLATION REQUIREMENTS
Install the isolator away from drafts and heat sources (radiators, ventilators/convectors), to ensure proper
functioning. For instance, in a small room (<30 m
3
), if an exhaust duct is used to expel air outside the
building, we recommend the installation of a grill in the room in order to provide an air supply equal at least
to the quantity entering in the isolator.
Install the isolator in a well-ventilated room with a low degree of dust.
The distance between the isolator and the exhaust system (if present) should be as short as possible.
Place the isolator away from doors and windows, which may cause malfunctions
Place the isolator in places where there is little human traffic.
The exhaust system should be installed on the roof or on a wall or through a window. Otherwise, it can be
connected to a chimney.
The door of the room should be in such a position relative to the isolator as to prevent drafts.
Min. temperature
5 °C
Max. temperature:
40 °C
Max. humidity:
80% at 31 °C, linear drop in relative humidity down to 50% relative humidity
at 40°C.
The exhaust duct (optional), located on top of the isolator, should be connected to the outside of the building,
alternatively it is possible the connection with the conditioning plant, but the air must not be recycled.
The maximum length of the
200mm exhaust hard duct should not exceed 10 linear meters. Otherwise
some changes must be made in the extraction fan or a remote exhaust motor-fan must be installed.
Before connecting the isolator to the mains power supply, check the necessary voltage and power indicated
on the plate near the power cable. The room must be equipped with an earth connection.
Содержание GLOVEFAST ASEPTIC
Страница 9: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 9 of 65...
Страница 26: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 26 of 65 1 2...
Страница 46: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 46 of 65...
Страница 48: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 48 of 65 15 J DIAGRAM FOR MAINTENANCE OPERATIONS...
Страница 49: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 49 of 65 16 FRONTAL DIAGRAM...
Страница 50: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 50 of 65 16 A SIDE DIAGRAM...
Страница 54: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 54 of 65 18 WIRING DIAGRAM...
Страница 55: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 55 of 65...
Страница 56: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 56 of 65...
Страница 57: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 57 of 65...
Страница 58: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 58 of 65...
Страница 59: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 59 of 65...
Страница 60: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 60 of 65...
Страница 61: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 61 of 65...
Страница 62: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 62 of 65...
Страница 63: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 63 of 65...
Страница 64: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 64 of 65...