5.C SYMBOLS OF THE CONTROL PANEL
List and description of all the symbols and controls of the touch screen control panel:
12
13
14
2
15
3
4
16
5
6
7
8
9
10
11
1
1 MAIN SWITCH:
Position "0"
in the "0" position, the green light of the mains voltage is on (3); the LCD
displays the model name "GLOVEFAST ASEPTIC ". In this position the
operator can activate only the lighting system (7), the optional U.V. lamp (14)
and the power outlet (6) (with plug installed) and can see the data stored in
the microprocessor by pressing the “STATUS” key (2).
Position "I"
by pressing "I" the password to enter is requested. When the password is
typed in (press arrow-
up key (▲) 5 times, arrow-down key (▼) 4 times) and
SET pressed the green led of the switch lights up and the isolator starts
operating, LAF and exhaust motor-blowers are powered and first "CHECK
PANEL" then "STAND-BY" appears on the display. The LCD shows a bar-
graph indicating the ongoing start-up procedure, where the laminar flow and
the internal pressure reach the pre-set values. In addition, an audible alarm
will sound intermittently during this stand-by period, alerting the operator not
to start working yet. When the audible alarm stops and the message
"STAND-BY" disappears from the display, the isolator is ready for use. The
LAF velocity and the internal pressure are displayed.
NOTE
: In any case, it is advisable to wait 5 minutes before starting to work.
2 DISPLAY
Backlight liquid crystal "LCD" display composed of 2 lines of 20 characters
each showing the operating parameters and alarms.
3 LINE
The green mains light switches on if the unit is connected to the mains and
the line is live.
4 WORKING CONDITION
The green LED lights up when the ventilation works correctly.
5 GAS
Not activated.
6 SOCKET
This
supplies
voltage; when enabled, the display shows "POWER ON" (total
current for all sockets: 4 Amp).
7 LIGHT
This switches on the lighting system; when enabled, the display shows
"Light on". Switching on the lighting system automatically the optional U.V.
lamp are switched off.
8 UP/DOWN ARROWS
Use the arrow keys to scroll the menu, to program changing parameters and
to put in the password. Three passwords are programmed: 1) to start the
Содержание GLOVEFAST ASEPTIC
Страница 9: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 9 of 65...
Страница 26: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 26 of 65 1 2...
Страница 46: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 46 of 65...
Страница 48: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 48 of 65 15 J DIAGRAM FOR MAINTENANCE OPERATIONS...
Страница 49: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 49 of 65 16 FRONTAL DIAGRAM...
Страница 50: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 50 of 65 16 A SIDE DIAGRAM...
Страница 54: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 54 of 65 18 WIRING DIAGRAM...
Страница 55: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 55 of 65...
Страница 56: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 56 of 65...
Страница 57: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 57 of 65...
Страница 58: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 58 of 65...
Страница 59: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 59 of 65...
Страница 60: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 60 of 65...
Страница 61: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 61 of 65...
Страница 62: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 62 of 65...
Страница 63: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 63 of 65...
Страница 64: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 64 of 65...