10.A Alarm List
Pos
English alarm message
German alarm message
Description
Setpoint
Tolerance Alarm limit
Time delay
1
Minimum LAF alarm
Alarm Minimum LAF
Airflow speed in the work chamber under minimum threshold value
0.45 m/s
+20%
-0,04 m/s from set point
1 sec
2
Maximum LAF alarm
Alarm max. LAF
Airflow speed in the work chamber over maximum threshold value
0.45 m/s
+20%
+0,04 m/s from set point
1 sec
3
Front glass open
Scheibe offen
The frontal glass is open.
n/a
n/a
n.a.
1 sec
4
Sensors failure
Fehler Sensor
Failure to one of the sensors. Press “MUTE” key and the
correspondent sensor is displayed
n/a
n/a
n.a.
1 sec
5
Black-out end
Stromausfall Ende
Warning of black-out when the cabinet is working. Press “MUTE” to
stop the alarm
n.a.
1 sec from the voltage
supply restoring
6
HEPA filters check
HEPA Filter pruefen
Possible clogging of absolute (HEPA) filters (The main motor power is
increased to the maximum in order to maintain the correct internal
pressure level)
n/a
n/a
97,5%
when the motor power
increase up to 97,5% (1
sec)
7
Exhaust duct check
Abluft Filt. Pruefen
Possible clogging of the exhaust duct or the isolator in not tight (the
exhaust motor power goes up to maximum in order to reach the
internal pressure set point)
n/a
n/a
100%
when the motor power
increase up to 100% (1
sec)
8
F% lifetime over
F% Betriebszeit Ende
Filter 1 (2, 3, 4, 5) residual lifetime is over
initial filters count
down is set to 10000
hours
n/a
n.a.
max 60sec (when the
minutes change in the
clock)
9
Min. PRESSURE Alarm
Alarm dp zu niedrig
The pressure of the work chamber is below the minimum limit. It can
be due to a hole in the gloves or to the opening of a door.
-40 Pa
0
-40Pa is the standard
alarm limit
4 seconds in normal
condition.
120 seconds if the hatch
doors are open
10
Open hatch panel
Schleusendeckel öffnen
No hatch panel is open. It is necesary to open them in order switch on
the ventilation of the isolator
n/a
n/a
n.a.
1 sec
LEGENDA:
Light blue cells
= it is possible to adjust the limit with the service password
Pos. 1, 2, 9
The alarm and the actual value are alternatiely displayed
Pos. 5
When there is a power failure the system go down and no alarm is displayed. When the power comes on the isolator will restore the last working conditions and the above described alarm will arise.
Pos. 6, 7, 8
Only alarm message is displayed; pressing "Status" button it is possible to check the actual value of the relative parameter
Pos. 3, 4, 10 Only alarm message is displayed
Содержание GLOVEFAST ASEPTIC
Страница 9: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 9 of 65...
Страница 26: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 26 of 65 1 2...
Страница 46: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 46 of 65...
Страница 48: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 48 of 65 15 J DIAGRAM FOR MAINTENANCE OPERATIONS...
Страница 49: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 49 of 65 16 FRONTAL DIAGRAM...
Страница 50: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 50 of 65 16 A SIDE DIAGRAM...
Страница 54: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 54 of 65 18 WIRING DIAGRAM...
Страница 55: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 55 of 65...
Страница 56: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 56 of 65...
Страница 57: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 57 of 65...
Страница 58: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 58 of 65...
Страница 59: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 59 of 65...
Страница 60: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 60 of 65...
Страница 61: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 61 of 65...
Страница 62: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 62 of 65...
Страница 63: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 63 of 65...
Страница 64: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 64 of 65...