INSTALLATION MANUAL - EN 

Edition  1.0  May 2021    

 

 

 

 

 

 

 

   Pag. 6 di 12 

 

2.4

 

COMPLETE LIGHT CONNECTION TO THE CEILING 

For the installation of such application please make reference to the dedicated sections of this manual. 
Connect the power supply cable according to the specifications outlined in Tab. 1. 

The applications are not supplied with the light. 
 

 

Warnings for electrical danger and suspended masses 

- The device has to be installed by specialized technicians only 
- The light has to be installed with FARO applications only. 

- The light is supplied with rotation limit switch between the fixed and the swivel  arm. 
- The switch limits must not be passed over or forced. 

 

2.4.1

 

Ceiling mounting with Faro applications 

 

Warnings for electrical danger and suspended masses 

The device must be installed by specialized tecnicians 

The power supply in the room where the fitting must be installed must be always switched off. 
Before starting the installation take care that the ceiling is suitable to bear the load of the application and light.  
The anchor bolts provided with the application must be used only with the following base materials: concrete, 
natural stone. 

Maximum load applicable for 60 seconds: 70 kg 
See tab 1 for the electrical requirements of the power supply.  
Stricly comply with the minimum heights shown below. 

 

 

 
 
 
 

 
 
 
 

Содержание MP3020

Страница 1: ......

Страница 2: ...2 3 3 Remote Cable connection 5 2 3 4 Auto on setting 5 2 4 COMPLETE LIGHT CONNECTION TO THE CEILING 6 2 4 1 Ceiling mounting with Faro applications 6 2 4 2 Installation procedure 7 2 4 3 Electrical d...

Страница 3: ...standard IEC 60364 1 and national installation regulations for electrical systems in premises for medical use When installed with ceiling floor or wall applications the device must be installed with a...

Страница 4: ...all the complete light Note 1 the fuse must be placed on the phase and not on the neutral The lamp must be installed with a multipolar device to separate it from the supply network meeting the require...

Страница 5: ...nsulation 1 85 mm Only use certified terminals and connectors with resistance to flame VW 1 or similar Take care that the device is protected with the fuse as recommended in Tab 1 2 3 3 Remote Cable c...

Страница 6: ...n limit switch between the fixed and the swivel arm The switch limits must not be passed over or forced 2 4 1 Ceiling mounting with Faro applications Warnings for electrical danger and suspended masse...

Страница 7: ...to the screws 8 and fix them with the nuts 9 and washers 10 G Calculate the proper height of the column 14 according to the Formula L H DHC 1020 mm Cut the exceeding column 14 part on the side were no...

Страница 8: ...on the column 14 and fix it with the three screws 19 make sure the bores of the bushing 20 match the screws seat on the column 14 and tighten Simultaneously pull the traction cord to push out the con...

Страница 9: ...ments of the power supply Stricly comply with the minimum heights shown below 1 Cap 2 Screw 3 Wall application 4 Wall plugs 5 Electrical connector 6 Connector cover 7 Carter 8 Screws 9 Washer 10 Cable...

Страница 10: ...cording to requirements of table 1 The resulting medical system must be declared in accordance with IEC EN 60601 1 by the installer Once the fixing point has been established with reference a the cent...

Страница 11: ...ch to screw the nut 4 Screw the 2 safety screws F until the cut A into the brass nut is completely closed 4 take care to leave free space around the nut G and the support S to avoid any friction when...

Страница 12: ...strenght 4000 V 2 MOPP between primary and sec Thermal Protection Component built in The medical system must be declared compliant with IEC EN60601 1 by the installation technician or manufacturer No...