INSTALLATION MANUAL - EN
Edition 1.0 May 2021
Pag. 3 di 12
1
SYMBOLS USED
1.1
SYMBOLS USED IN THE MANUAL
WARNING
The paragraphs marked with this symbol contain instructions that must be carefully followed to avoid damaging the
device, harming the operator or the patient.
CAUTION
These instructions warns you that you must pay attention to avoid situations that could damage the device.
FORBIDDEN
This icon highlights what you should not do to avoid damaging the device.
NOTES
This icon supplies information that allows you to use the device more efficiently.
1.2
GENERAL STANDARDS AND MAIN WARNINGS
General Warning
The device can be installed in a medical system (unit), but also on specific applications to ceiling, floor or wall.
The device can be powered both by the unit and by a power supply connected directly to the mains. See the
specific installation paragraph.
The device must never be modified without written authorization of FARO S.p.A.
For maintenance only original FARO’s spare parts are allowed. Fail to comply with this warning will immediately
make decay the warranty and the conformity of the product to international regulations and directives on
Medical Devices.
Warnings against electrical danger and fire
Installation of the device must only be carried out by qualified personnel.
The light must be installed on a specific control and power supply device, such as units, or with an electrical
system that meets standard IEC 60364-1 and “national installation regulations for electrical systems in premises for
medical use”.
When installed with ceiling, floor or wall applications the device must be installed with an omnipolar separation
device from the mains and compliant with Standard IEC 61058-1. This separation device must be approved to
withstand 4 kV of transient voltage.
Installation and maintenance of device conformity with the standard IEC 60601-1 is the responsibility of the
installation technician or the manufacturer of the medical system / unit).
Check that the power supply voltage, indicated on the main plate, corresponds to the mains voltage.
Warning for danger of explosion
The device is not suitable for installation in environments with the presence of flammable gas or rich of oxygen.
1.3
STORAGE AND USE: ENVIRONMENTAL PROVISIONS
The device must be used in the following environmental conditions:
• Temperature from 10° to 40°C
• Max altitude: 2000 m
• Relative humidity from 30% to 75%
2
DEVICE INSTALLATION
Warnings for electrical danger
The device must be installed by specialist technicians.
On installation, the power supply must always be disconnected.
Refer to the wiring diagrams in the manual.
Check the main plate data before installation
Note for installation
The power supply cable on the complete light is supplied without any connector or terminal to allow connection
according to the specifications of the combination or application.
The functionality and safety of the light does not depend on the polarity of the power supply current. Therefore
inversion of the power supply cables will not pose a risk of malfunctioning.