INSTALLATION MANUAL - EN 

Edition  1.0  May 2021    

 

 

 

 

 

 

 

   Pag. 3 di 12 

 

1

 

SYMBOLS USED 

1.1

 

SYMBOLS USED IN THE MANUAL 

 

WARNING 

The paragraphs marked with this symbol contain instructions that must be carefully followed to avoid damaging the 

device, harming the operator or the patient. 

 

CAUTION 

These instructions warns you that you must pay attention to avoid situations that could damage the device. 

 

FORBIDDEN 

This icon highlights what you should not do to avoid damaging the device. 

 

NOTES 

This icon supplies information that allows you to use the device more efficiently. 
 

1.2

 

GENERAL STANDARDS AND MAIN WARNINGS 

 

 

General Warning 

 

The device can be installed in a medical system (unit), but also on specific applications to ceiling, floor or wall. 
The device can be powered both by the unit and by a power supply connected directly to the mains. See the 
specific installation paragraph. 
The device must never be modified without written authorization of FARO S.p.A.  

For maintenance only original FARO’s spare parts are allowed. Fail to comply with this warning will immediately 
make  decay  the  warranty  and  the  conformity  of  the  product  to  international  regulations  and  directives  on 
Medical Devices.

 

 

Warnings against electrical danger and fire 

Installation of the device must only be carried out by qualified personnel. 
The light must be installed on a specific control and power supply device, such as units, or with an electrical 

system that meets standard IEC 60364-1 and “national installation regulations for electrical systems in premises for 
medical use”.  
When installed with ceiling, floor or wall applications the device must be installed with an omnipolar separation 
device from the mains and compliant with Standard IEC 61058-1. This separation device must be approved to 

withstand 4 kV of transient voltage. 
Installation and maintenance of device conformity with the standard IEC 60601-1 is the responsibility of the 
installation technician or the manufacturer of the medical system / unit).  
Check that the power supply voltage, indicated on the main plate, corresponds to the mains voltage.  

 

Warning for danger of explosion 

The device is not suitable for installation in environments with the presence of flammable gas or rich of oxygen. 

 

1.3

 

STORAGE AND USE: ENVIRONMENTAL PROVISIONS 

The device must be used in the following environmental conditions: 
• Temperature from 10° to 40°C 
• Max altitude: 2000 m 

• Relative humidity from 30% to 75% 
 
 

2

 

DEVICE INSTALLATION 

 

Warnings for electrical danger 

The device must be installed by specialist technicians. 

On installation, the power supply must always be disconnected.  
Refer to the wiring diagrams in the manual. 
Check the main plate data before installation 

 

Note for installation 

The power supply cable on the complete light is supplied without any connector or terminal to allow connection 

according to the specifications of the combination or application. 
The functionality and safety of the light does not depend on the polarity of the power supply current. Therefore 
inversion of the power supply cables will not pose a risk of malfunctioning. 

Содержание MP3020

Страница 1: ......

Страница 2: ...2 3 3 Remote Cable connection 5 2 3 4 Auto on setting 5 2 4 COMPLETE LIGHT CONNECTION TO THE CEILING 6 2 4 1 Ceiling mounting with Faro applications 6 2 4 2 Installation procedure 7 2 4 3 Electrical d...

Страница 3: ...standard IEC 60364 1 and national installation regulations for electrical systems in premises for medical use When installed with ceiling floor or wall applications the device must be installed with a...

Страница 4: ...all the complete light Note 1 the fuse must be placed on the phase and not on the neutral The lamp must be installed with a multipolar device to separate it from the supply network meeting the require...

Страница 5: ...nsulation 1 85 mm Only use certified terminals and connectors with resistance to flame VW 1 or similar Take care that the device is protected with the fuse as recommended in Tab 1 2 3 3 Remote Cable c...

Страница 6: ...n limit switch between the fixed and the swivel arm The switch limits must not be passed over or forced 2 4 1 Ceiling mounting with Faro applications Warnings for electrical danger and suspended masse...

Страница 7: ...to the screws 8 and fix them with the nuts 9 and washers 10 G Calculate the proper height of the column 14 according to the Formula L H DHC 1020 mm Cut the exceeding column 14 part on the side were no...

Страница 8: ...on the column 14 and fix it with the three screws 19 make sure the bores of the bushing 20 match the screws seat on the column 14 and tighten Simultaneously pull the traction cord to push out the con...

Страница 9: ...ments of the power supply Stricly comply with the minimum heights shown below 1 Cap 2 Screw 3 Wall application 4 Wall plugs 5 Electrical connector 6 Connector cover 7 Carter 8 Screws 9 Washer 10 Cable...

Страница 10: ...cording to requirements of table 1 The resulting medical system must be declared in accordance with IEC EN 60601 1 by the installer Once the fixing point has been established with reference a the cent...

Страница 11: ...ch to screw the nut 4 Screw the 2 safety screws F until the cut A into the brass nut is completely closed 4 take care to leave free space around the nut G and the support S to avoid any friction when...

Страница 12: ...strenght 4000 V 2 MOPP between primary and sec Thermal Protection Component built in The medical system must be declared compliant with IEC EN60601 1 by the installation technician or manufacturer No...

Отзывы: