MP3020
INSTALLATION MANUAL - EN
Edition 1.0 May 2021
Pag. 12 di 12
2.8.2
Electrical Requirements
The requirements for correct installation of the head are as follows:
Power Supply Supply cables
Power Supply requirements
Type
of
Device
Conformity
to
IEC
60601-1
17-24 Vac
50/60 Hz
Cable:
2 cables UL Style
1061
RED
300 V
T 80°C
1x26 AWG
VW 1
Ø max 1,02mm
Connector:
Molex 43645-0200
Transformer complies with IEC/EN 60601-
1 third edition.
Protection on secondary circuit with at
least one appropriate fuse:
T1.6AL250V
Minimum requirements:
Output: 17 - 24 Vac;
Minimum Power: 26 VA;
Dielectric strenght > 4000 V;
2 MOPP between primary and sec.
Thermal Protection
Component
built-in
The medical
system must
be declared
compliant with
IEC/EN60601-1 by the
installation technician
or
manufacturer.
Note for the
installation
technician: ensure
the combined version
on which the light is
installed is certified to
host the complete
light.
25-32 Vdc
Power supply conform to IEC/EN 60601-1
third edition with one pole protected by
appropriate fuse:
T1.25AL250V
Minimum requirements:
Output: 32 Vdc;
Power: min 14 VA;
Dielectric strenght > 4000 V;
2 MOPP between primary and sec.
Continuous protection from short
circuit or overcurrent
Component
built-in
Tab 2 – Requirements for electrical connection and compliance with IEC/EN 60601-1)
3
ELECTRICAL DRAWINGS
D : Connector for Remote Cable (option)