27
EN
1.3 PURPOSE
Name:
TRACHEOFIRST
®
COMPACT
Main function:
Temporary and spontaneous aspiration of secretions, blood
and body fluids that typically arise during suctioning of the
airways.
Medical indication / application:
Suctioning of the upper airways
Specification of the main functions:
Drainage and temporary collection of body secretions. A
vacuum is generated by means of an electric suctioning pump.
The integrated secretion container enables temporary collection
of the body secretions that have been drained off.
Application organ:
Upper airways (oral cavity, nasopharyngeal space and
bronchial system)
Duration of application:
Temporary application in patients undergoing suctioning of the
airways (operating time of the battery up to 60 minutes)
Application environment:
The application environment for airway suctioning is the clinical,
medical practice, nursing and homecare field. Application must
only be carried out by medically trained and briefed persons.
Contraindication:
Not suitable for:
● continuous operation in connection with drainage in the
low-vacuum range (e.g. thoracic drainage or wound
drainage).
● long-lasting use in connection with endoscopy.
● use outside of the medical field.
● aspiration of flammable, caustic or explosive substances.
● suctioning in areas exposed to explosion hazards.
TRACHEOFIRST
®
COMPACT is an active
medical device.
Reprocessing:
The product is intended for repeated use. The accessories are
partly reusable. For information on reprocessing, cleaning and
disinfection, refer to instructions for use (see Section 6.0).
TRACHEOFIRST
®
COMPACT must not be used:
•
for continuous operation in connection with drainage in the low-vacuum range
(e.g. thoracic drainage) or long-lasting use in connection with endoscopy;
•
outside of the medical field (for aspiration of flammable, caustic or explosive liquids/gases)
N
Figure 1
TRACHEOFIRST
®
COMPACT
➊
Vacuum gauge
➋ Vacuum control
➌ On / Off button
➍ Bacterial filter
➎ Connection for power supply unit
➏ Secretion container
➐ Connecting tube
➑ Secretion outlet with stopper
➒ Connection for suction tube with stopperl
1.2 FUNCTION
●
TRACHEOFIRST
®
COMPACT is a particularly handy
small-size suctioning device.
● The device is powered by a maintenance-free electric swing-
piston pump. When in operation, the pump generates a
vacuum in the tubing system and in the secretion container,
with the aid of which secretions or rinsing fluids are suctioned
off (via a sterile suction catheter).
● The fluid is collected in the secretion container.
● The final vacuum, and thus also the suction power, can be
adjusted with the aid of the fine control and the vacuum
gauge.
● A bacterial filter prevents intrusion of bacteria into the inside
of the unit.
● In addition, a mechanical overflow protector (float) integrated
on the lid of the secretion container prevents accidental
aspiration of secretions into the pump head.
1.0 INTRODUCTION
➊
➍
➋
➌
➐
➏
➎
➑
➒