44
EN
12.0 INSTRUCTIONS WITH REGARD TO EMC
The TRACHEOFIRST
®
COMPACT device must not be used in an arrangement where it is placed directly next to other devices
or stacked with other devices. If the device has to be operated close to or stacked with other devices, the device should be ob
-
served in order to verify its proper use when used in this arrangement
12.2 GUIDELINES AND MANUFACTURER'S DECLARATION -
ELECTROMAGNETIC IMMUNITY FOR TRACHEOFIRST
®
COMPACT
TRACHEOFIRST
®
COMPACT is intended for operation in the electromagnetic environment specified below. The customer or
user of the TRACHEOFIRST
®
COMPACT device should ensure that the device is used in such an environment.
Electromagnetic
immunity tests
IEC 60601- test level
Accordance level
Electromagnetic environment -
guidelines
Electrostatic discharge
(ESD) as per IEC 61000-
4-2
± 6 kV contact discharge
± 8 kV air discharge
± 6 kV contact discharge
± 8 kV air discharge
Floors should be made of wood or
concrete or tiled with ceramic tiles. If the
floor is covered with synthetic material,
the relative air humidity must be at least
30 % betragen.
Fast transient electric dis-
turbances/bursts as
per IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input and output
lines
± 2 kV for power supply
lines
The quality of the supply voltage should
be equivalent to that of a typical home or
hospital environment.
Surge voltages
(surges) as per
IEC 61000-4-5
± 1 kV common mode
voltage
± 2 kV series mode voltage
± 1 kV common mode
voltage
The quality of the supply voltage should
be equivalent to that of a typical home or
hospital environment.
Voltage dips, short
interruptions and variations
of the power supply voltage
as per IEC 61000-4-11
< 5 % U
T
(> 95 % dip of U
T
for
0.5 cycle)
40 % U
T
(60% dip of U
T
for 5 cycles)
70 % U
T
(30% dip of U
T
for 25 cy-
cles)
< 5 % U
T
(> 95 % dip of U
T
for 5 s)
< 5 % U
T
(> 95 % dip of U
T
for
0.5 cycle)
40 % U
T
(60% dip of U
T
for 5 cycles)
70 % U
T
(30% dip of U
T
for 25 cy-
cles)
< 5 % UT
(> 95 % dip of U
T
for 5 s)
The quality of the supply voltage should
be equivalent to that of a typical home
or hospital environment. If the
TRA-
CHEOFIRST
®
COMPACT user requires
continued functions even when the
power supply is interrupted, it is recom-
mended to power the TRACHEOFIRST
®
COMPACT from an uninterruptible power
supply or a battery.
Magnetic field at the supply
frequency (50/60 Hz) as
per IEC 61000-4-8
3 A/m
30 A/m
Magnetic fields at the mains power sup
-
ply frequency should be equivalent to the
typical values found in the business and
hospital environment.
NOTE U
T
is the alternating mains power supply voltage before application of the test levels.
Содержание TRACHEOFIRST COMPACT
Страница 1: ...TRACHEOFIRST COMPACT GEBRAUCHSANWEISUNG INSTRUCTIONS FOR USE ...
Страница 2: ......
Страница 47: ......