![EZ Digital DS-1000 Series Скачать руководство пользователя страница 45](http://html.mh-extra.com/html/ez-digital/ds-1000-series/ds-1000-series_operation-manual_2457878045.webp)
Rising Time :
This displays a rising time of currently displayed waveforms.
When range of input signal is out of the ADC input range, peak to peak value is
less than 2 division, or width of 1 cycle is less than 0.4 division on display, "?" is
displayed. when 1 cycle is not captured, "invalid" is displayed.
Falling Time :
This displays a falling time of currently displayed waveforms.
When range of input signal is out of the ADC input range, peak to peak value is
less than 2 division, or width of 1 cycle is less than 0.4 division on display, "?" is
displayed. When 1 cycle is not captured, "Invalid" is displayed.
Period
:
This displays a period of 1cycle of currently displayed waveforms.
When range of input signal is out of the ADC input range, peak to peak value is
less than 2 division, or width of 1 cycle is less than 0.4 division on display.
When 1 cycle is not captured, "Invalid" is displayed.
+ Width
:
This displays a positive width of 1 cycle of currently displayed
waveforms. When range of input signal is out of the ADC input range, peak to
peak value is less than 2 division, or width of 1 cycle is less than 0.4 division on
display, "?" is displayed. When 1 cycle is not captured, "Invalid" is displayed.
- Width
:
This displays a negative width of 1 cycle of currently displayed
waveforms of channel. When range of input signal is out of the ADC input range,
peak to peak value is less than 2 division, or width of 1 cycle is not captured,
"Invalid" is displayed.
Duty
:
This displays a duty ratio of 1 cycle of currently displayed waveforms
of channel. When range of input signal is out of the ADC input range, peak to
peak value is less than 2 division, or width of 1 cycle is less than 0.4 division on
display, "?" is displayed. When 1 cycle is not captured, "Invalid" is displayed.
44
Содержание DS-1000 Series
Страница 1: ...Digital Oscilloscope ...
Страница 2: ...Digital Oscilloscope ...
Страница 8: ...1 INTRODUCTION 7 ...
Страница 22: ...21 ...
Страница 23: ...2 DESCRIPTION 22 ...
Страница 25: ...10 CH1 CH2 POSITION CURSOR 11 TRIGGER LEVEL CURSOR 12 HORIZONTAL TRIGGER POSITION 13 INPUT COUPLING MODE 24 ...
Страница 61: ...3 APPENDIX 60 ...
Страница 67: ...The specifications are subjected to change without notice MMA282 620A ...