1-3-3 PROBE
The first step of measurement is to connect the signals to the instrument
properly.
BODY
HOOK COVER
RETRACTABLE
COVER
GROUND CLIP
CAPACITANCE
TRIMMER
CORRECTION
GROUND
HOOK TIP
Probe Grounding
Connect the probe ground lead as close as possible to the point being
measured especially when measuring a signal with a fast rise time or a
high frequency signal. Long probe ground leads may cause waveform
distortions, such as ringing and overshoot.
Attenuation switch
CH 1
Spring GND
When the optional probe is used with the x10 / x1 select switch set to x10,
the input signal to the instrument is attenuated to 1/10. When a signal is
too small to be measured with x10, use the mode x1. In this case, note
that the input impedance of x1 is different from that of x10, and the mea-
surable frequency band becomes very low.
Probe Attenuation Setting
18