4-2. Installation Conditions
4-1. Installation Environment
30 cm
30cm
Air Inlet
Air Outlet
Model FDU-1110: Approx. 100 kg
Model FDU-2110: Approx. 140 kg
Air Inlet
(Slits)
- 14 -
Installation
4
Use extreme caution about installation
location, air-conditioning and ventilation
Since this product uses an air-cooled-type refrigerator,
heat is discharged from the air outlet. Due to this
feature, not to raise ambient temperature, provide
sufficient ventilation system for the product.
Otherwise, use the product in air-conditioned space (in
which ambient temp.: 5 to 35
C) appropriately.
If ambient temperature is raised, operation efficiency
may be deteriorated, and cooling capacity may be
dropped too.
Furthermore, operation of refrigerator under hot
temperature and high pressure may result in equipment
failure.
WARNING
Keep enough space around the product
CAUTION
Be careful in carrying the heavy product
CAUTION
If the product is installed in inappropriate environment,
It may result in equipment damage in an early stage,
or, insufficient exercise of its functions
and performance.
Be careful of installation environment
CAUTION
Install this product in places such as:
●
Not subject to the direct sunlight.
●
Room temperature can be maintained within 5 to
35
°
C.
●
Well-vented, or venting is ensured.
●
No flammable or ignitable substances, liquids, and
gases placed around.
●
No condensation.
●
Low humidity and no splash of water.
●
Little dusts.
●
Horizontal, stable and solid.
(Check the weight of the product during the operation.)
To keep the product’s performance, be sure
to secure the space between product and
wall or ceiling, etc. as shown in the figure on
the right.