- 13 -
Alarm Messages
Causes of Alarms
Alarm Operations
Displays
Alarm Functions
Guidewire of temp. sensor
has malfunction. *1
Detection of
malfunction
in guidewire
of temp. sensor
Lit
Alarm lamp is lighted, and “0000”
is indicated on the Trap Temp.
Display and the buzzer is sounded
(for 10 seconds).
Control is continued in manual
operation. In Auto operation,
the mode is switched into Manual
operation, and control is continued.
0000
Trap Temp. Display
PrEr
Vacuum Gauge Display
Lit
*1: To release an alarm message, clear the cause of alert, and then press [Set] key. The display returns to
normal indication.
*2: Turn on the power to reset the alarm.
Alarm
Alarm
Power failure
alarm
Vacuum
Error Alarm
PUPm
Vacuum Gauge Display
Lit
Control Board
Error Alarm
Total control is stopped, and
the buzzer is sounded continuously.
(Due to abnormal status, no
message is specified)
Automatic recovery is attempted.
Message
not specified
Alarm
Alarm
oFF
Current trap temp.
and “OFF”
are indicated in turn.
-
Measured
temp.
Backup Battery
Error Alarm
CPAr
Vacuum Gauge Display
Lit
Alarm lamp is lighted, and “CPAr”
is indicated on Trap Temp. Display.
Then the buzzer is sounded
(for 10 seconds), and total control is
stopped.
* Elapsed time for model 1 is cleared.
* In case of power failure, total
control is stopped despite preset
period for operation recovery
time in power failure.
Alarm
Alarm of
malfunction
in guidewire
of vacuum sensor
Alarm lamp is lighted, and “PrEr”
is indicated on the vacuum gauge
display. Then the buzzer is sounded
(for 10 seconds).
Control is continued in manual
operation. In Auto operation,
the mode is switched into Manual
operation, and control is continued.
Circuit of vacuum Sensor
has malfunction. *1
Alarm lamp is lighted, and current
trap temp. and “OFF” are indicated
in turn. Then the buzzer is sounded
(for 10 seconds).
The power recovers according to
preset power-failure-recovery
settings that had been selected.
The power failed during control
,or, the power was turned off
without terminating control. *1
Alarm lamp is lighted, and “PUPm”
is indicated on the vacuum gauge
display. Then the buzzer is sounded
(for 10 seconds).
Vacuum pump is stopped solely
in Manual operation.
In Auto operation, the pump is
stopped solely, other units are
switched into manual operation,
and their control are continued.
Vacuum range of 533 Pa or more
is indicated on vacuum gauge
display once. And “Atmo”
is indicated for 30 min. or more.
*1
Backup Battery failed or runs
out
* Press [Set] key to release alarm
message. By doing this, normal
indication recovers. However,
backup function for recovery
select for power failure and
elapsed time for Model 1 do not
operate normally.
Due to noise interference or
others, abnormal status occurred
to the control Board. Automatic
recovery cannot be executed
*2