background image

 

DM220   V3.0   01/06 

Diode Test

 

Test current 1mA, open circuit voltage 1.5V DC typical 

Continuity Check

 

Audible sound if the resistance is less than 100

 

Non-Contact Volt detect 

100 to 600VAC; 50/60Hz 

Input Impedance

 >7.5M

 (VDC & VAC) 

AC Response 

Average responding 

ACV Bandwidth 

50Hz to 60Hz 

Display 

4000 count 3-3/4 digit liquid crystal 

Overrange indication

  “OL” is displayed 

Polarity

 

Automatic (no indication for positive); Minus (-) sign for 

negative 

Auto Power Off 

15 minutes (approx) 

Low Battery Indication  “    ”

 is displayed if battery voltage drops below operating 

voltage 

Batteries

 

Two 1.5 volt AAA

 

Fuses

  

mA, µA ranges; 0.2A/500V fast acting resettable fuse 

Operating Temperature

 32ºF to 104ºF (0ºC to 40ºC) 

Storage Temperature

 14

o

F to 122

o

F (-10

o

C to 50

o

C) 

Operating Humidity 

Max 80% up to 87ºF (31ºC) decreasing linearly to 50% at 

104ºF (40ºC) 

Storage Humidity 

<80%  

Operating Altitude 

7000ft. (2000meters) maximum. 

Weight

  

0.319lb (145g) 

Size

  

4.09” x 2.1” x 1.2” (104 x 55 x 32.5mm) 

Safety 

For origin of installation use and in accordance with the 
requirements for double insulation per EN61010-1 and 
IEC61010-1 2

nd

 Edition (2001) to Overvoltage Category IV 

600V and Category III 1000V; Pollution Degree 2. 

Warranty 

EXTECH INSTRUMENTS CORPORATION

 warrants this instrument to be free of defects in parts 

and workmanship for 

three years 

from date of shipment (a six month limited warranty applies to 

sensors and cables). If it should become necessary to return the instrument for service during or 

beyond the warranty period, contact the Customer Service Department at (781) 890-7440 ext. 210 

for authorization or visit our website www.extech.com for contact information. A Return 

Authorization (RA) number must be issued before any product is returned to Extech. The sender is 

responsible for shipping charges, freight, insurance and proper packaging to prevent damage in 

transit. This warranty does not apply to defects resulting from action of the user such as misuse, 

improper wiring, operation outside of specification, improper maintenance or repair, or 

unauthorized modification. Extech specifically disclaims any implied warranties or merchantability 

or fitness for a specific purpose and will not be liable for any direct, indirect, incidental or 

consequential damages. Extech's total liability is limited to repair or replacement of the product. 

The warranty set forth above is inclusive and no other warranty, whether written or oral, is 
expressed or implied. 

 

 

 
 
 
 
 
 
 

Copyright 

©

 2006 Extech Instruments Corporation 

All rights reserved including the right of reproduction in whole or in part in any form. 

Support line (781) 890-7440 

Technical support: Extension 200; E-mail: [email protected] 

Repair & Returns: Extension 210; E-mail: [email protected] 

Product specifications subject to change without notice 

For up-to-the-minute product information, visit our website: www.extech.com 

Extech Instruments Corporation, 285 Bear Hill Rd., Waltham, MA 02451

 

www.burntec.com

Содержание DM220

Страница 1: ...he Extech DM220 Pocket Autoranging Multimeter This meter measures AC DC Voltage AC DC Current with 200mA 500V resettable fuse Resistance Capacitance Frequency Duty Cycle Diode Test and Continuity It features a built in non contact AC voltage detector plus flashlight Proper use and care of this meter will provide many years of reliable service www burntec com ...

Страница 2: ...ode Test Continuity Test 600V DC AC 2 USE EXTREME CAUTION when working with voltages greater than 25VAC or 35VDC These voltages are considered a shock hazard 3 DO NOT measure voltage if the voltage on the COM input jack exceeds 600V above earth ground 4 NEVER connect the meter leads across a voltage source while the function switch is in the current resistance or diode mode Doing so can damage the...

Страница 3: ...t probe tip to the positive side of the circuit 4 Read the voltage in the display AC DC CURRENT MEASUREMENTS 1 For current measurements up to 4000µA AC DC set the function switch to the µA position 2 For current measurements up to 200mA AC DC set the function switch to the mA position 3 Press the MODE button to indicate AC or DC on the display 4 Remove power from the circuit under test then open u...

Страница 4: ...n the display 3 Touch the test leads to the capacitor to be tested 4 Read the capacitance value in the display FREQUENCY MEASUREMENTS 1 Set the rotary function switch to the V Hz position 2 Press the Hz button to indicate Hz on the display 3 Touch the test probe tips to the circuit under test 4 Read the frequency on the display DUTY CYCLE 1 Set the rotary function switch to the V Hz position 2 Pre...

Страница 5: ...CLEAN Wipe the case occasionally with a damp cloth DO NOT use chemicals cleaning solvents or detergents 5 USE ONLY FRESH BATTERIES OF THE RECOMMENDED SIZE AND TYPE Remove old or weak batteries so they do not leak and damage the unit 6 IF THE METER IS TO BE STORED FOR A LONG PERIOD OF TIME the batteries should be removed to prevent damage to the unit BATTERY INSTALLATION WARNING To avoid electric s...

Страница 6: ...igits 40kΩ 0 01kΩ 400kΩ 0 1kΩ 0 8 reading 2 digits 4MΩ 0 001MΩ 3 0 reading 5digits Resistance 40MΩ 0 01MΩ 5 0 reading 5digits 4nF 0 001nF 5 0 reading 0 6nF 40nF 0 01nF 3 5 reading 30 digits 400nF 0 1nF 4µF 0 001µF 3 0 reading 20 digits 40µF 0 01µF Capacitance 200µF 0 1µF 5 0 reading 30 digits 9 999Hz 0 001Hz 99 999Hz 0 001Hz 999 9Hz 0 1Hz Frequency 9 999kHz 0 001kHz 1 0 reading 2 digits Duty Cycle...

Страница 7: ...ty applies to sensors and cables If it should become necessary to return the instrument for service during or beyond the warranty period contact the Customer Service Department at 781 890 7440 ext 210 for authorization or visit our website www extech com for contact information A Return Authorization RA number must be issued before any product is returned to Extech The sender is responsible for sh...

Страница 8: ...DM220 V3 0 01 06 8 www burntec com ...

Страница 9: ...ímetro de escala automática modelo DM220 de Extech Este medidor lee voltaje CA CD corriente CA CD con fusible de reposición de 200mA 500V resistencia capacitancia frecuencia ciclo de trabajo prueba de diodo y continuidad Ofrece un detector integrado de voltaje CA sin contacto más linterna El uso y cuidado apropiado de este medidor le proveerá muchos años de servicio confiable www burntec com ...

Страница 10: ...iodo prueba de continuidad 600V CD CA 2 EXTREME SUS PRECAUCIONES al trabajar con voltajes mayores a 25VCA o 35VDC Estos voltajes son considerados un peligro de choque 3 NO mida voltajes si el voltaje en el enchufe de entrada COM excede 600V sobre tierra física 4 NUNCA conecte los cables del medidor a una fuente de voltaje cuando el selector de función esté en modo de corriente resistencia o diodo ...

Страница 11: ...to Toque la punta de la sonda roja de prueba del lado positivo del circuito 4 Lea el voltaje en la pantalla MEDICIÓN DE CORRIENTE CD CA 1 Para medidas de corriente hasta 4000µA CA CD fije el selector de función en la posición µA 2 Para medidas de corriente hasta 200mA CA CD fije el selector de función en la posición mA 3 Presione el botón MODE para indicar CA o CD en la pantalla 4 Corte la tensión...

Страница 12: ...r 4 Lea el valor de capacitancia en la pantalla MEDIDAS DE FRECUENCIA 1 Fije el selector giratorio de función en la posiciónV Hz 2 Presione el botón Hz para indicar Hz en la pantalla 3 Toque las puntas de las sondas a través del circuito bajo prueba 4 Lea la frecuencia en la pantalla CICLO DE TRABAJO 1 Fije el selector giratorio de función en la posición V Hz 2 Presione el botón Hz para selecciona...

Страница 13: ...NGA LIMPIO EL MEDIDOR Ocasionalmente limpie la caja con un paño húmedo NO use químicos solventes para limpieza o detergentes 5 USE SÓLO BATERÍAS NUEVAS DEL TAMAÑO Y TIPO RECOMENDADO Retire las baterías viejas o débiles de manera que no se derramen y dañen la unidad 6 SI SE VA A ALMACENAR EL MEDIDOR DURANTE UN LARGO PERIODO DE TIEMPO deberá retirar la batería para prevenir daños a la unidad Instala...

Страница 14: ...kΩ 0 01kΩ 400kΩ 0 1kΩ 0 8 lectura 5 dígitos 4 MΩ 0 001MΩ 3 0 lectura 5 dígitos Resistencia 40MΩ 0 01MΩ 5 0 lectura 5 dígitos 4 nF 0 001nF 5 0 lectura 0 6 nF 40nF 0 01nF 3 5 lectura 30 dígitos 40nF 0 1nF 4µF 0 001µF 3 0 lectura 20dígitos 40µF 0 01µF Capacitancia 200µF 0 1µF 5 0 lectura 30 dígitos 9 999Hz 0 001Hz 99 999Hz 0 001Hz 999 9Hz 0 1Hz Frecuencia 9 999kHz 0 001kHz 1 0 lectura 2 dígitos Ciclo...

Страница 15: ...ables y sensores Si fuera necesario regresar el instrumento para servicio durante o después del periodo de garantía llame al Departamento de Servicio a Clientes al teléfono 781 890 7440 ext 210 para autorización o visite nuestra página en Internet en www extech com para Información de contacto Se debe otorgar un número de Autorización de Retorno RA antes de regresar cualquier producto a Extech El ...

Отзывы: