5                                      CO210-en-EU_V1.2   8/14 

P1.

3

 CO

2

 ALARM: HIGH ALARM

 

P1.

3

 is used for setting the CO

2

 upper limit for the BEEPER ALARM level. The current set value 

will be blinking on the display. 
Press the 

LOG/

or 

MIN/MAX

 button to increase or decrease the value. Each press adjusts in 

100ppm increments. The alarm range is 1000 to 5000ppm. 
When the value has been set, press the 

SET

 button to confirm the limit and to proceed to P2.0

 or 

p

ress the 

ESC

 button to return to P1.

0

 
 
 
 
 

 
P2.

0

 TEMPERATURE SCALE 

Press the

 LOG/

 button in P1.

0

 mode to access P2.

0

 for setting the temperature scale.  

Press the 

SET

 button to go into the P2.1setting mode. The °C or °F will blink.  

Press the 

LOG/

button to change the units. Press the SET button to confirm  the setting or press 

the 

ESC

 button to exit without saving and return to P2.

0

 
 
 
 
 

P3.

0

 REAL TIME CLOCK 

Press the 

LOG/

 button in P1.

0

 twice to access P3.

0

 for setting up real time clock. Press the 

SET

 

button and the meter goes into P3.

1

 with blinking year in the lower left display. To change the year, 

press the 

LOG/

 button or the 

MIN/MAX

 button. Press the 

SET 

button to save the setting and 

then enter P3.

2

 or press the 

ESC

 button return to P3.

0

 without saving the setting. 

Press the 

LOG/

 button P3.

1

 to access P3.

2

. The current month setting will blink. To change the 

month, press the 

LOG/

 button or the 

MIN/MAX

 button. Press the 

SET

 button to save the setting 

and then enter P3.

3

 or press the 

ESC

 button return to P3.

0

 without saving the setting. 

Repeat as above to finish the setting of P3.

3

 (Date), P3.

4

 (Hour) and P3.

5

 (Minute) 

 
 
 
 
 
 
 

CO

2

Содержание 793950502105

Страница 1: ...User s Guide CO2 Monitor and Datalogger Model CO210...

Страница 2: ...Date and Time 3 Air Temperature 4 Relative Humidity Symbols ppm CO2 value GOOD CO2 air quality level NORMAL CO2 l air quality level POOR CO2 air quality level Air Temp Air Temperature Humidity Relati...

Страница 3: ...he RESET button for more than 1 second to clear the minimum and maximum value from memory DATA LOGGING The meter can record readings of CO2 temperature and humidity for long time environment monitorin...

Страница 4: ...ode for more than 1 second to enter the setup mode P1 1 CO2 ALARM GOOD LEVEL When entering the setup mode P1 0 and CO2 are displayed on the LCD Press the SET button again to go into P1 1 for setting t...

Страница 5: ...ange the units Press the SET button to confirm the setting or press the ESC button to exit without saving and return to P2 0 P3 0 REAL TIME CLOCK Press the LOG button in P1 0 twice to access P3 0 for...

Страница 6: ...settings Parameter Default P1 1 700ppm P1 2 1000ppm P1 3 1000ppm P2 1 C P4 1 No P5 0 SAMPLING RATE Press the LOG button in P1 0 three times to access P5 0 to set the logging sample rate The available...

Страница 7: ...ation is suggested to be done outdoors on a sunny day with good ventilation and fresh air where the CO2 level is approximately 400 ppm Do not calibrate on a rainy day because high humidity will affect...

Страница 8: ...Humidity 0 1 to 99 9 0 1 3 10 to 90 5 10 or 90 Display LCD Sensor Type CO2 NDIR non dispersive infrared technology Humidity Capacitance sensor Temperature air Thermistor Response CO2 2min for 90 step...

Страница 9: ...if it still appear re calibrate the meter If above two methods faile return for repair E17 ABC mode of CO2 senor has failed and might cause wrong CO2 readings Send back for repair Temp Display E02 Air...

Страница 10: ...lead to serious oxygen deprivation resulting in permanent brain damage coma and even death Regulatory exposure limits ASHRAE Standard 62 1989 1000ppm CO2 concentration in occupied building should not...

Отзывы: