8 CO210-en-EU_V1.2 8/14
Maintenance
CLEANING AND STORAGE
1. The meter should be cleaned with a damp cloth and mild detergent when necessary. Do not use
solvents or abrasives.
2. Store the meter in an area with moderate temperature and humidity.
Specifications
Function Range
Resolution
Accuracy
CO
2
0 to 9999ppm
1ppm
±(5%rdg +50ppm)
Temperature
-10 to 60°C
14 to 140°F
0.1° ±0.6°C/0.9°F
Humidity
0.1 to 99.9%
0.1%
±3%(10 to 90%)
±5%(< 10% or > 90%)
Display LCD
Sensor Type
CO
2
: NDIR (non-dispersive infrared) technology
Humidity: Capacitance sensor;
Temperature (air): Thermistor
Response CO
2
: <2min for 90% step change
Temp: <2min for 90% step change
%RH: <10min for 90% step change
Datalogging
Up to 15999 points
Sample rate
1 second to 4 hours-59 minutes-59 seconds
Operating Conditions
-10 to 60°C (14 to 140°F); < 90% RH non-condensing
Storage Conditions
-20 to 60°C (-4 to 140°F); <99% RH non-condensing
Power Supply
5VDC (±10%),
≥
500mA (AC adaptor supplied)
Dimensions / Weight
117x102x102mm (4.6x4x4”); 204g (7.2 oz.)