7                                      CO210-en-EU_V1.2   8/14 

Calibration 

CO

2

 CALIBRATION 

The meter is calibrated to a standard 400ppm CO

2

 concentration at the factory  

NOTE:

 When the accuracy becomes a concern or after a year of use, return the meter to Extech for 

calibration. 

CAUTION:

 Do not calibrate the meter in an atmosphere of unknown CO

2

 concentration.  

 

ABC (Automatic Baseline Calibration)

 

ABC (Automatic Baseline Calibration) establishes a baseline calibration to eliminate the zero drift of 
the infrared sensor. The ABC function is always “ON” when the meter is turned on. ABC is designed 
to calibrate the meter at the minimum CO

2

 reading detected during 7.5 days continuous monitoring 

(power on). It assumes that the area being tested receives fresh air with a CO

2

 level of 

approximately 400ppm at some period of time during the seven days. 

It is not suitable to use 

desktop CO

2

 in closed areas with consistently high CO

2

 levels 24 hours a day

Manual Calibration

 

Manual calibration is suggested to be done outdoors on a sunny day with good ventilation and fresh 
air where the CO

2

 level is approximately 400 ppm. Do not calibrate on a rainy day because high 

humidity will affect the CO

2

 level in air.  

 
Do not calibrate in places crowded with people or close to where high CO

2

 concentrations may exist 

such as ventilating outlets or fireplaces. 
 
Place the meter in the calibration site. Turn on the meter and hold down the 

SET

 and 

MIN/MAX

 buttons simultaneously for more than 1 second to enter CO

2

 calibration mode. 

“400ppm” and “CO

2

” will blink while the calibration is in process. 

 
 
 
 
 
 
The calibration will take about 30 minutes. When the calibration is complete, the blinking stops and 
the meter returns to normal operation. To abort the calibration, press the RESET button for more 
than 1 second. 

Note: 

Keep away from any animal, human or plant which might affect the CO

2

 concentration during 

the calibration. 

Software 

The Windows based pc software provided is used to set-up the datalogger, download data and to 
view the data in graphical or text format. After installing the software, view the HELP file for 
operational information. 

Содержание 793950502105

Страница 1: ...User s Guide CO2 Monitor and Datalogger Model CO210...

Страница 2: ...Date and Time 3 Air Temperature 4 Relative Humidity Symbols ppm CO2 value GOOD CO2 air quality level NORMAL CO2 l air quality level POOR CO2 air quality level Air Temp Air Temperature Humidity Relati...

Страница 3: ...he RESET button for more than 1 second to clear the minimum and maximum value from memory DATA LOGGING The meter can record readings of CO2 temperature and humidity for long time environment monitorin...

Страница 4: ...ode for more than 1 second to enter the setup mode P1 1 CO2 ALARM GOOD LEVEL When entering the setup mode P1 0 and CO2 are displayed on the LCD Press the SET button again to go into P1 1 for setting t...

Страница 5: ...ange the units Press the SET button to confirm the setting or press the ESC button to exit without saving and return to P2 0 P3 0 REAL TIME CLOCK Press the LOG button in P1 0 twice to access P3 0 for...

Страница 6: ...settings Parameter Default P1 1 700ppm P1 2 1000ppm P1 3 1000ppm P2 1 C P4 1 No P5 0 SAMPLING RATE Press the LOG button in P1 0 three times to access P5 0 to set the logging sample rate The available...

Страница 7: ...ation is suggested to be done outdoors on a sunny day with good ventilation and fresh air where the CO2 level is approximately 400 ppm Do not calibrate on a rainy day because high humidity will affect...

Страница 8: ...Humidity 0 1 to 99 9 0 1 3 10 to 90 5 10 or 90 Display LCD Sensor Type CO2 NDIR non dispersive infrared technology Humidity Capacitance sensor Temperature air Thermistor Response CO2 2min for 90 step...

Страница 9: ...if it still appear re calibrate the meter If above two methods faile return for repair E17 ABC mode of CO2 senor has failed and might cause wrong CO2 readings Send back for repair Temp Display E02 Air...

Страница 10: ...lead to serious oxygen deprivation resulting in permanent brain damage coma and even death Regulatory exposure limits ASHRAE Standard 62 1989 1000ppm CO2 concentration in occupied building should not...

Отзывы: