84
If the Interlock contacts are opened during a test, the pop-up message will be displayed and
the test will abort. The hardware and has been configured to provide the interlock
connections on pins 4 and 5 of the Remote Interface, Signal Input port. The instrument can
still be used
If the Interlock contacts are opened during a test, the pop-up message will be displayed and
the test will abort. The hardware and has been configured to provide the interlock
connections on pins 4 and 5 of the Remote Interface, Signal Input port. The instrument can
still be used without the external interlock device as long as the Interlock Connector (1505
provided with unit) is plugged into the Remote Interface, Signal Input port. If there is
nothing connected to the Remote Interface, Signal Input port to provide a connection to the
interlock, the instrument will not perform tests.
PLC Remote Pop-up message
If you attempt to start a test from the front panel Test button and the PLC remote function is
turned “ON”, a pop-up message will be displayed. The pop-up message will appear as
follows:
P L C r e m o t e
C o n t r o l i s O N .
F r o n t p a n e l T E S T b u t t o n i s
d i s a b l e d
.