45
run the currently loaded test file.
Disabling the password
To disable the password simply change the password to “0”. After the password is reset to
“0” all security functions may be access without the use of a password.
Forgotten Password
If you have forgotten your password you may access all security functions by typing in the
number “8000” into the password field. The old password can not be recovered. A new
password should be entered or enter “0” to disable the password.
4.1.5 Power-On Screen
The Power-On screen allows you to select the screen(s) that will be displayed when the
instrument is first powered up. From the System Setup screen, press the “Power-On
Screens” soft key. The Power-On screen will now be displayed. The Power-On screen will
appear as follows:
From the Power-On screen, two different parameters may be accessed, 7630 Info and Main
Menu.
4.1.5.1 7630 Info
The 7630 Information screen is the very first screen to appear when the instrument is powered
up. You have the option to PAUSE the instrument at this screen after power up or
CONTINUE through to the next screen after the Information screen is displayed for four
seconds. The option to PAUSE or CONTINUE may be selected by pressing the 7630 Info
soft key. When PAUSE is selected the words “Press any key to continue“ will appear at the
bottom of the 7630 Information screen.
4.1.5.2 Main Menu
When the Main Menu parameter is ON, the first display to appear after the 7630 Information
screen is the Main Menu. When the Main Menu parameter is OFF the first display to appear
after the 7630 Information screen is the Perform Test screen.