104
Connect 0.125VDC and a voltmeter to Probe-HI + and Probe-LO -. When the power supply
and meter are connected, press TEST to activate the calibration. Read the measurement from
your standard and enter it using the numeric keypad. You may now store the new calibration
number by pressing the ENTER key or escape by pressing the EXIT key or the RESET button.
13.
LLT
×
256 DC range
:
Calibration of LLT MD Volts DC x 256
Connect 0.032VDC and a voltmeter to Probe-HI + and Probe-LO -. When the power supply
and meter are connected, press TEST to activate the calibration. Read the measurement from
your standard and enter it using the numeric keypad. You may now store the new calibration
number by pressing the ENTER key or escape by pressing the EXIT key or the RESET button.
14.
LLT
×
0.1 AC range :
Calibration of LLT MD Volts AC x 0.1