53
multiple parameter allows for set up of the scanner channels. The individual scanner channel
parameters may be accessed using the arrow keys and the state of the scanner channel may be
selected using a softkey labeled “Scanner Channel Output Select”. The three different
selectable scanner states are L (scanner channel set to the return point), H (scanner set to the
high voltage point) and O (OFF).
Defaults
: Pressing this softkey will default all of the test parameters to the predetermined
default values. Once the default soft key is pressed, it is necessary to complete the edit either
by pressing the ENTER key to accept the parameter overwrite or the EXIT key to escape from
the edit and return to the original values.
4.3.2 Run Test (OPT.752)
From the Setup Tests, Tests Selection screen, press the “Run Test” soft key. The Run Test
parameter-setting screen will now be displayed. The Run test parameter-setting screen will
appear as follows:
From the Run Test parameter-setting screen, the following parameters may be controlled:
Voltage-HI, Voltage-LO, Amp-HI, Amp -LO, Delay Time, Dwell Time, Leakage-HI,
Leakage-LO, Power-HI, Power-LO, PF-HI, PF-LO and
Continuous
.
An additional parameter exclusive for DUT input power is Power Control.
4.3.2.1 Voltage Setting (Voltage-HI and Voltage-LO)
Please use the up and down arrow keys or the ENTER key and scroll the highlighted area to
the
Voltage-HI
or
Voltage-LO
parameter. The display will appear as follows:
Voltage-HI Voltage-LO