54
Enter the new voltage via the numeric keypad and then press the ENTER key to accept the
new parameter or press the EXIT key to escape from the edit.
4.3.2.2 Current Setting (Amp-HI and Amp-LO)
Please use the up and down arrow keys or the ENTER key and scroll the highlighted area to
the
Amp-HI
or
Amp-LO
parameter. The display will appear as follows:
Amp -HI Amp -LO
Enter the new current via the numeric keypad and then press the ENTER key to accept the
new parameter or press the EXIT key to escape from the edit.
4.3.2.3 Delay Time and Dwell Time setting
Please use the up and down arrow keys or the ENTER key and scroll the highlighted area to
the
Delay Time
or
Dwell Time
parameter. The display will appear as follows:
Delay Time Dwell Time
Enter the new time via the numeric keypad and then press the ENTER key to accept the new
parameter or press the EXIT key to escape from the edit.
If the dwell time is set to “0”, the instrument will operate in a continuous mode. It will stop
when the DUT (Device Under Test) goes into failure or the “RESET” button is pressed. The
minimum dwell time setting is 0.1 seconds.