Rear Panel Controls (Continuing)
14. POWER CONTROL BUS (OPTION)
This is only for OPT.752 RUN Test Function. The power source must be connected to this
Connector, then it can supply power to DUT by 7128.
15. SCANNER OUTPUT
This is an I/O signal control bus. There are 7 signals are provided by seven normally open
internal relays. They can to used to control the HI/LO signals of Probe HI and Probe LO of
DUT by Scanner.
16. MD OUTPUT SIGNAL TERMINAL
With external voltage meter or oscillograph it is convenient for verifying the circuit of MD.
Optional OPT.754 High Measurement Range 35mArms / 70mApeak & 4MDS may be
substituted for built-in 7 kinds of common measuring devices
。
NOTE
:
Must connect with isolated oscilloscope to BNC connect to take waveform
17. EXTERNAL MEASURING DEVICE
This compartment contains an external Measuring Device PCB. This device will enable
during a Touch Current test when the “External” is selected from the Meas. Device soft key.
The external measuring device allows the operator to test to a standard that calls for a custom
measuring device or easily change the measuring device if the standard requirements change.
The instrument is provided with a 1k ohm resistor on the measuring device PCB. The PCB
can easily be reconfigured for a simple resistive component or a complex two-pole measuring
device.