380193-en-GB_V2.6 8/16
9
SPECIFICATIONS
Capacitance @ 120Hz
Range
Cx accuracy
DF accuracy
Note
9.999mF
±(5.0% rdg + 5d)
(DF<0.1)
±(10%rdg + 100/Cx + 5d)
(DF<0.1)
after short cal
1999.9
μ
F
±(1.0% rdg + 5d)
(DF<0.1)
±(2%rdg + 100/Cx + 5d)
(DF<0.1)
after short cal
199.99
μ
F
±(0.7% rdg + 3d)
(DF<0.5)
±(0.7%rdg + 100/Cx + 5d)
(DF<0.1)
19.999
μ
F
±(0.7% rdg + 3d)
(DF<0.5)
±(0.7%rdg + 100/Cx + 5d)
(DF<0.1)
1999.9nF
±(0.7% rdg + 3d)
(DF<0.5)
±(0.7%rdg + 100/Cx + 5d)
(DF<0.1)
199.99nF
±(0.7% rdg + 5d)
(DF<0.5)
±(0.7%rdg + 100/Cx + 5d)
(DF<0.5)
after open cal
19.999nF
±(1.0% rdg + 5d)
(DF<0.1)
±(2.0%rdg + 100/Cx + 5d)
(DF<0.1)
after open cal
Capacitance @ 1kHz
Range
Cx accuracy
DF accuracy
Note
999.9
μ
F
±(5.0% rdg + 5d)
(DF<0.1)
±(10%rdg + 100/Cx + 5d)
(DF<0.1)
after short cal
199.99
μ
F
±(1.0% rdg + 3d)
(DF<0.5)
±(2.0%rdg + 100/Cx + 5d)
(DF<0.5)
after short cal
19.999
μ
F
±(0.7% rdg + 3d)
(DF<0.5)
±(0.7%rdg + 100/Cx + 5d)
(DF<0.1)
1999.9nF
±(0.7% rdg + 3d)
(DF<0.5)
±(0.7%rdg + 100/Cx + 5d)
(DF<0.1)
199.99nF
±(0.7% rdg + 5d)
(DF<0.5)
±(0.7%rdg + 100/Cx + 5d)
(DF<0.1)
19.999nF
±(0.7% rdg + 5d)
(DF<0.1)
±(0.7%rdg + 100/Cx + 5d)
(DF<0.1)
after open cal
1999.9pF
±(1.0% rdg + 5d)
(DF<0.1)
±(2.0%rdg + 100/Cx + 5d)
(DF<0.1)
after open cal
Inductance @ 120Hz
Range
Lx accuracy (DF<0.5)
DF accuracy (DF<0.5)
Note
10000H
Not specified
Not specified
1999.9H
±(1.0%rdg + Lx/10000 + 5d)
±(2.0%rdg + 100/Lx + 5d)
after open cal
199.99H
±(0.7%rdg + Lx/10000 + 5d)
±(1.2%rdg + 100/Lx + 5d)
19.999H
±(0.7%rdg + Lx/10000 + 5d)
±(1.2%rdg + 100/Lx + 5d)
1999.9mH
±(0.7%rdg + Lx/10000 + 5d)
±(1.2%rdg + 100/Lx + 5d)
199.99mH
±(1.0%rdg + Lx/10000 + 5d)
±(3.0%rdg + 100/Lx + 5d)
after short cal
19.999mH
±(2.0%rdg + Lx/10000 + 5d)
±(10%rdg + 100/Lx + 5d)
after short cal
Inductance @ 1kHz
Range
Lx accuracy (DF<0.5)
DF accuracy (DF<0.5)
Note
1999.9H
Not specified
Not specified
199.99H
±(1.0%rdg + Lx/10000 + 5d)
±(1.2%rdg + 100/Lx + 5d)
after open cal
19.999H
±(0.7%rdg + Lx/10000 + 5d)
±(1.2%rdg + 100/Lx + 5d)
1999.9mH
±(0.7%rdg + Lx/10000 + 5d)
±(1.2%rdg + 100/Lx + 5d)
199.99mH
±(0.7%rdg + Lx/10000 + 5d)
±(1.2%rdg + 100/Lx + 5d)
19.999mH
±(1.2%rdg + Lx/10000 + 5d)
±(5.0%rdg + 100/Lx + 5d)
after short cal
1999.9
μ
H
±(2.0%rdg + Lx/10000 + 5d)
±(10%rdg + 100/Lx + 5d)
after short cal
Note
: Where Lx or Cx is the the C or L reading in the display without range indication.
i.e. For a reading of 18.888, use 18888 as the factor.